FISCHER Mess- und Regeltechnik GMBH
Installation and assembly | 3
BA_EN_ME11
11/24
b) Three-wire connection
Pressure Transmitter
Supply
Output signal
Input signal
+Sig
+/~U
b
-/~U
b
-Sig
+/~U
b
-/~U
b
AC/DC
AC/DC
AC
Fig. 10:
3L circuitry AC/DC
3.3.1 Standardised plug DIN EN 175 301-803-A
1
3
2
Cable screw connection
M16x1.5
Seal NBR
Contact insert
Housing
Screw
1
2
3
Line socket
Flanged connector
Fig. 11:
Line socket DIN EN 175 301-803-A
Ter-
minal
Signal name
DC
Cable colour
1
Supply /Output
+U
b
+Sig red
2
Supply /Output
-U
b
-Sig
blue
3
n.c.
n.c.
Table 1:
2-wire connection 4 ... 20 mA
Ter-
minal
Signal name
AC
DC
Cable colour
1
Output
+Sig Black
2
Supply /Output
~U
b
-U
b
-Sig
blue
3
Supply
~U
b
+U
b
red
n.c.
Table 2:
3-wire connection 0 ... 10V
The earth connection in the standardized plug is not connected.
Содержание ME11 Series
Страница 1: ...Operating manual ME11 Pressure Transmitter 09005184 BA_EN_ME11 Rev ST4 B 04 22 09005184...
Страница 21: ...FISCHER Mess und Regeltechnik GMBH Attachments 7 BA_EN_ME11 21 24 7 Attachments Fig 16 CE_EN_ME11...
Страница 22: ...7 Attachments FISCHER Mess und Regeltechnik GMBH 22 24 BA_EN_ME11 Fig 17 UKCA_EN_ME11...
Страница 23: ...FISCHER Mess und Regeltechnik GMBH BA_EN_ME11 23 24 Notes...