11  Technical data 

Basic measuring range  

mbar 

10 

16 

25  40  60  100  ±2.5 

±4 

±6 

±10 

±16 

±25  ±40  ±60  ±100 

Pa 

400  600  1000  1600 

 

 

 

 

±250 

 

 

 

 

 

 

 

 

kPa 

 

 

1.6  2.5 

10 

 

 

 

±1 

±1.6  ±2.5  ±4 

±6 

 

Max. stat. operating pres-
sure 

mbar 

50 

100 

250 

500 

50 

100 

250 

500 

Bursting pressure 

mbar 

150 

300 

750 

1500 

150 

300 

750 

1500 

Characteristic 
curve deviation °) 

max. 

%FS 

1.0 

1.0 

typ. 

%FS 

0.5 

0.5 

TC span°°) 

max. 

%FS/10K 

1.0 

0.3 

1.0 

0.5 

0.3 

typ. 

%FS/10K 

0.3 

0.3 

TC zero point °°) 

max. 

%FS/10K 

1.0 

0.4 

1.0 

0.5 

0.4 

typ. 

%FS/10K 

0.2 

0.2 

°:  Characteristic  curve  deviation  (non-linearity  and  hysteresis)  at  25°C,  basic  measuring  range  (linear  characteristic  curve,  not  spread) 
°°)  :  in relation to basic measuring range (not spread), compensation range 0 ... 60°C 

 

 

General points 

 

Admissible ambient temperature 

-10 ... 70°C 

 

Admissible media temperature 

-10 ... 70°C 

 

Admissible storage temperature 

-20 ... 70°C 

 

Enclosure protection class 

IP 65 acc. to DIN EN 60529 

 

 

Electrical data 

 

Nominal voltage 

24 V DC / AC 

 

Admissible operating voltage U

b

 

12 ... 32 V DC / AC 

 

Electrical connection type 

Three-wire 

 

Output signal 

Current output 

Voltage output 

 

 

0 ... 20 mA, 4 ... 20 mA 

0 ... 10 V DC 

 

Admissible apparent ohmic 

resistance 

R

L

 

≤ (U

b

 - 4 V) / 0.02 A  

(U

b

 

≤ 26V) 

R

L

 

≤1100Ω  

(U

b

 > 26V) 

R

L

 

≥   2 KΩ  

(U

b

 

≥ 15V) 

R

L

 

≥ 10 KΩ  

(U

b

 

= 12 …15V) 

 

Characteristic curve 

can be programmed 

 

Power consumption 

approx. 2 W / VA 

 

Display 

3.5 character LED 

 

 

Switch contacts 

 

Relay contacts 

 

2 potential-free relay contacts programmed as NO contact or NC contact 
U

max

 = 32V AC/DC, I

max

 = 2 A, P

max

 = 64 W/VA 

 

Semiconductor switch 

2 potential-free semiconductor switches (MOSFET), SPST-NO/NC progr. 
U = 3 … 32V AC/DC, I

max

 = 0.25A, P

max

 = 8 W/VA, R

ON

 

≤ 4 Ω 

 

 

Connections 

 

Process connection 

Hose screw connections made of Al, 6/4 mm or 8/6 mm 

 

electr. connection 

 
 

2 x round plug connector M12 
Connector 1 for supply and analogue output signal (5-pole, male) 
Connector 2 for switch contacts (4-pin, male) 

 

 

Materials 

 

Casing 

Polyamide PA 6.6 

 

Media-contacting material 

Silicon, PVC, aluminium, brass 

 

 

Assembly 

 

 

Bore-holes on the reverse side for attachment of the assembly panels. 
Wall mounting by means of assembly plate. 
If the device is intended for outdoor use, we recommend permanently protecting the 
membrane keypad against UV radiation and using a suitable enclosure or at least the erection 
of a sufficiently dimensioned canopy as a protection measure against constant rain or snow. 

Positive ranges (0 … ) 

+/- ranges 

Содержание DE44

Страница 1: ...zed personnel familiar with the in stallation commissioning and operation of this product Specialized personnel are persons who can assess the work they have been assigned and recognize potential dang...

Страница 2: ...eed between the manufacturer and user 3 Description of the product and func tional description 3 1 Function diagram 3 2 Design and mode of operation The device is based on two piezo resistive sensor e...

Страница 3: ...ilters need to be installed at suitable points The pressure sensing lines need to be kept as short as possible and installed without sharp bends to avoid interfering delay times If the pressure sensin...

Страница 4: ...tions step by step Depending on the device model 1 some menu items may not be available For example all characteristic curve functions are faded out in the menu if the de vice does not have a signal o...

Страница 5: ...isplay output signal and switch points but not on the measuring cell it self This parameter can be used to set the reaction time to the pressure jumps The value range is 0 0 s to 100 0 s But with maxi...

Страница 6: ...means of hydrostatic pressure measure ments are two typical examples of this In these cases you might want to change the output signal of the transmitter to a non linear characteristic curve so that t...

Страница 7: ...ue 000 cancels the password function If a password has been issued the text PAS appears after ESc and and you need to enter the correct value using and You will only arrive at all other menu items aft...

Страница 8: ...e of the de vice can be shown sensibly MA Start of measuring range The measuring value is set in that the output signal is minimal e g 0V 0mA or 4mA ME End of measuring range The measuring value is se...

Страница 9: ...must not be exposed to mechanical shocks It shall be transported only in packaging specifically intended for transport 8 Service All damaged or faulty devices shall be directly sent to our repair depa...

Страница 10: ...4 20 mA 0 10 V DC Admissible apparent ohmic resistance RL Ub 4 V 0 02 A Ub 26V RL 1100 Ub 26V RL 2 K Ub 15V RL 10 K Ub 12 15V Characteristic curve can be programmed Power consumption approx 2 W VA Di...

Страница 11: ...ith 3 30 support points Password 001 999 000 no password protection Comments 0 For values from 0 2 the display switches between dP1 and dP2 rhythmically 1 Measuring values around zero are set to zero...

Страница 12: ...5 6 0 40 mbar 5 7 0 60 mbar 5 8 0 100 mbar 5 9 2 5 2 5 mbar A 6 4 4 mbar A 7 6 6 mbar A 8 10 10 mbar A 9 16 16 mbar B 1 25 25 mbar B 2 40 40 mbar C 5 60 60 mbar B 3 100 100 mbar B 4 0 400 Pa D 7 0 60...

Страница 13: ...40 mbar 5 7 0 60 mbar 5 8 0 100 mbar 5 9 2 5 2 5 mbar A 6 4 4 mbar A 7 6 6 mbar A 8 10 10 mbar A 9 16 16 mbar B 1 25 25 mbar B 2 40 40 mbar C 5 60 60 mbar B 3 100 100 mbar B 4 0 400 Pa D 7 0 600 Pa D...

Страница 14: ...sured value display contact elements 3 1 2 digit LED 2 relay contacts 3 3 1 2 digit LED 2 semiconductor switches 6 Electrical connection M12 plug connection M Assembly option Standard attachment boreh...

Страница 15: ...14 Declaration of Conformity...

Страница 16: ...ject to change without notice Changements techniques sous r serve Fischer Mess und Regeltechnik GmbH Bielefelder Str 37a D 32107 Bad Salzuflen Tel 49 5222 9740 Fax 49 5222 7170 eMail info fischermesst...

Отзывы: