background image

01

06

02

07

03

08

04

09

05

10

a rate of 1-3 inches (2.5-7.5cm) per second.

3  Once the SCL2 indicates a change in concen-

tration (audible beep), note the spot on the 

line and keep the tip moving past the potential 

leak to refresh the air space within the SCL2 

with clean air. (Note: If the SCL2 wand moves 

through a very high concentration of refriger-

ant, you may need to use it in fresh air for 4 

seconds before moving to step 4 in order to 

clear the refrigerant within the SCL2.)

4  Return the tip back across the spot first in-

dicated. When the SCL2 indicates a second 

change, note the spot on the line. The source 

of the leak will be near the midpoint between 

the two noted spots of indication. 

5  Very high concentrations of refrigerant can 

cause an overload which can take a few sec-

onds to clear.

 Quick Start

1  Charge the SCL2 (see charging instructions 

on page 14 for details).

2  Power on SCL2 and allow a 30 second warm 

up to complete in well ventilated ambient air 

away from suspected leak.

3  Select the sensitivity using the L/M/H button.

4  Point the tip of the wand at suspected leak 

locations (see Find CO2 Leaks section on page 

4 for details).

Quick Tips

1  Keep the wand tip moving slowly past sus-

pected leak locations.

2  Once a leak is detected, sweep the wand back 

over to pinpoint.

3  To pinpoint larger leaks adjust sensitivity.

Certifications 

C-Tick (N22675)

CE

WEEE

RoHS Compliant

Meets SAE Standard

Description

The SCL2 is a portable CO2 (R-744) leak 

detector with a superior combination of 

sensitivity, speed, sensor life, battery life, 

portability, and ease of use.

Find leaks on commercial refrigeration 

systems, restaurant soda machines, or any other 

system using R-744. Pinpoint tiny leaks in high 

concentration CO2 environments with ease.

Gas is pumped through the tip of the wand 

into the sensor within the SCL2 body. The sensor 

detects changes in concentration, not absolute 

concentration, allowing the SCL2 to differentiate 

CO2 concentrations higher than concentrations 

normally found in ambient air.

The SCL2 detects leaks as small as 6 grams/

yr (0.2 oz/yr). The SCL2 has three sensitivity 

settings to keep false triggering to a minimum. 

A quick and automatic 30 second self-calibration 

upon power-up ensures optimal performance. 

The SCL2 comes with a wall and car charger 

for its ultra-compact lithium-ion battery 

which powers the SCL2 for 8hrs of continuous 

operation before a recharge is needed. That's 

long enough to last your entire workday.

A built-in replaceable filter blocks moisture 

and harmful particulates. 

Find CO2 Leaks

Keep the SCL2 away from any areas of 

potential refrigerant leakage until the warm-up 

and calibration period is over. The warm-up and 

calibration sequence lasts about 30 seconds 

after powered on. Allow the SCL2 to fully warm-

up before searching for CO2 leaks to avoid false 

triggering. The meter defaults to high sensitivity.

The most likely points for refrigerant leaks 

are at the soldered joints in refrigerant lines 

and changes in the cross section or direction 

of these lines.  

The SCL2 detects changes in concentration of 

CO2, not the absolute concentration. This helps 

you to detect CO2 leaks in places that may have 

lots of CO2, such as a refrigerator leaking into 

an enclosed space. Because you are detecting 

a change in concentration, you should use the 

"double-pass" method detailed below. 

1  The tip of the wand should be close to the line. 

You may need to be within 1/4” (0.63cm) of 

a small leak to detect it. In this case, using a 

second hand to guide the tip along refrigerant 

lines may be helpful. See Figure 1.

2  Keep the tip moving along refrigerant lines at 

Fieldpiece

Carbon
Dioxide
Leak
Detector

OPERATOR'S
MANUAL

Model SCL2

Buttons

ON/OFF Protection

To turn on/off the SCL2 press and hold 

SCL2

Carbon Dioxide

Leak Detector

SENSITIVITY

H
M
L

PEAK

LOW-BATT

PRESS FOR

1 SECOND

BATTERY

CHECK

ON/OFF

MUTE

PEAK

L/M/H

for one second. This slight delay protects against 

your meter turning on when stored. If you forget 

to turn it off it will turn off automatically in 10 

minutes.

L/M/H Sensitivity

Set the sensitivity level by pressing 

SCL2

Carbon Dioxide

Leak Detector

SENSITIVITY

H
M
L

PEAK

LOW-BATT

PRESS FOR

1 SECOND

BATTERY

CHECK

ON/OFF

MUTE

PEAK

L/M/H

. Low(L), medium(M), 

or high(H) sensitivity will be indicated by their respective LED. 

The higher the concentration of refrigerant in the ambient air, 

the lower the sensitivity setting should be to minimize false trips.

Mute and Battery Check

Pressing 

SCL2

Carbon Dioxide

Leak Detector

SENSITIVITY

H
M
L

PEAK

LOW-BATT

PRESS FOR

1 SECOND

BATTERY

CHECK

ON/OFF

MUTE

PEAK

L/M/H

 toggles the sound of the SCL2 

on and off. In addition, holding the MUTE button 

for one second will display the remaining 

percent of battery charge on the LED bar graph.

Peak Mode

To activate Peak mode, press 

SCL2

Carbon Dioxide

Leak Detector

SENSITIVITY

H
M
L

PEAK

LOW-BATT

PRESS FOR

1 SECOND

BATTERY

CHECK

ON/OFF

MUTE

PEAK

L/M/H

. When 

activated, the highest LED will stay lit when 

gas is detected. The Peak function holds the 

highest change in concentration sensed while 

continuing to detect leaks. This is useful when 

detecting around corners when the LED bar 

graph is not visible. To deacviate, press PEAK 

button again.  The last peak  measured will be 

cleared.

Turbo Mode

The TURBO function is an additional 4th 

sensitivity that pushes the SCL2 to it highest 

possible sensitivity. This feature is activated by 

pressing 

SCL2

Carbon Dioxide

Leak Detector

SENSITIVITY

H
M
L

PEAK

LOW-BATT

PRESS FOR

1 SECOND

BATTERY

CHECK

ON/OFF

MUTE

PEAK

L/M/H

 four times in succession, and can 

only be activated while on the high(H) sensitivity 

level. 

When in TURBO mode the first green LED on 

SCL2 will strobe, and the audio detection sound 

will be a solid tone. 

To exit the TURBO mode, press the PEAK 

button again four times in succession, or switch 

from high(H) to one of the other sensitivities 

levels.

Beware: While in TURBO mode the SCL2 is 

extremely sensitive and will more easily trigger 

on abrupt or violent motions as well as very 

small changes in refrigerant levels.  Handle with 

care when in TURBO mode.

LED Indicators

LED Bar Graph Display

The eight segment LED display indicates the degree of change 

in refrigerant concentration. As the concentration of refrigerant 
in the air increases,  so does the number of lit bars on the display.

Low-Batt LED

1  While detector is in use, the Low-Batt LED will turn solid red 

indicating approximately 1 hour of battery life left.

2  While detector is charging, the Low-Batt LED will flash red on 

and off indicating the meter is charging.

3  When detector is fully charged, the Low-Batt LED will turn off 

indicating the meter is fully charged.

H/M/L LEDs

High (red), medium (amber) , or low (green) 

LED will turn solid indicating what sensitivity 

is selected.

Peak LED

Light solid amber when Peak mode is 

activated.

SCL2

Carbon Dioxide

Leak Detector

SENSITIVITY

H
M
L

PEAK

LOW-BATT

PRESS FOR

1 SECOND

BATTERY

CHECK

ON/OFF

MUTE

PEAK

L/M/H

SCL2

Carbon Dioxide

Leak Detector

SENSITIVITY

H
M
L

PEAK

LOW-BATT

PRESS FOR

1 SECOND

BATTERY

CHECK

ON/OFF

MUTE

PEAK

L/M/H

SCL2

Carbon Dioxide

Leak Detector

SENSITIVITY

H
M
L

PEAK

LOW-BATT

PRESS FOR

1 SECOND

BATTERY

CHECK

ON/OFF

MUTE

PEAK

L/M/H

!

  WARNING 

!

DO NOT operate the SCL2 in the presence 

of gasoline, natural gas, propane, or in other 

combustive atmospheres.

DO NOT use the SCL2 without the proper 

filter correctly installed.

99 Washington Street 
Melrose, MA 02176 

Phone

 781-665-

1400

Toll Free 1-800-517-8431

Visit us at www.T

estEquipmentDepot.com

Отзывы: