3  ABOUT THE PGX+ 

3.1 

Instrument specifications

Camera

Image Capture Rate: 

80 frames/second

Field of View: 

7,0 x 5,2 mm

Sensor Resolution:  

640 x 480 pixels

Pixel resolution:   

11 µm

Droplet application

Droplet Volume: 

0,1 – 10 µl

Modes:  

Touch Down / Impact / Manual

Communication: 

USB 2.0 

Operating conditions

Temperature: 

10 - 35 °C

Humidity: 

20 - 70%

Dimensions

Length: 

90 mm

 

Width:

55 

mm

 

Height:

90 

mm

Net. Weight: 

± 0.5 kg

3.2 

PC specifications

Minimum specifications

1GB of RAM
20MB of free disk space
1024x768 screen resolution
USB port

Recommended system

Core i5 5th generation Intel (or compatible) processor
8GB of RAM
100MB of free disk space
1920x1080 screen resolution
USB port
Laser/Inkjet Printer

Operating systems

Windows Vista, Windows 7, Windows 8, Windows 8.1 or Windows 10

3.3 

Scope of supply

Your PGX+ instrument is delivered in a carrying case (see picture) including

PGX+ software

one millilitre syringe

calibration plate

glass bottle

PGX+ Accessory kit with syringe tip, pump tubing and applicator 
tubing

NOTE:           
Carefully check all small wrappings, boxes and envelopes in the large 
corrugated box to be sure all parts and supplies, as listed on the packing list, 
are present.

3.4 

Consumables and accessoiries

Art. No. 

Description

 

PG1023 

Liquid Container Kit (with & w/o plug), 5+5 pcs

PG1171 

Steel cannulus tip 0.5 mm (min order quantity 10 pcs)

PG1025 

Applicator Tubing between Syringe and Tubing Dispenser, 5 pcs

PG1021 

Pump Tubing (PTFE) suitable for most liquids, 5 pcs

PG1027 

Pump Tubing (Silicone) suitable for Diiodomethane, 2 pcs

PG1010 

Dosing unit

Calibration plate  

PGX Instrument

Syringe  

Accessory Kit  

Glass bottle

1               2

3               4  

5

Содержание 68-76-00-0001

Страница 1: ......

Страница 2: ......

Страница 3: ...ng the eventual transport protections Before shipping an authorization to return should be secured from Fibro System by telephone FAX or written letter The best way to ship will also be advised at tha...

Страница 4: ...4...

Страница 5: ...ial specimens 13 6 3 Pump dispenser 14 7 Software functions 15 7 1 Graphmaster camera 15 7 2 Graphmaster test 16 7 3 Graphmaster database 20 8 Test methods 21 8 1 Static Mode 21 8 2 Dynamic mode 22 8...

Страница 6: ...6...

Страница 7: ......

Страница 8: ...ing the instrument is connected to the special PC software The software gives you an easy overview of the possible tests During the test the data transferred to the computer The pictures taken and the...

Страница 9: ...sta Windows 7 Windows 8 Windows 8 1 or Windows 10 3 3 Scope of supply Your PGX instrument is delivered in a carrying case see picture including PGX software one millilitre syringe calibration plate gl...

Страница 10: ......

Страница 11: ...and I follow the displayed instructions When the installation is terminated the new device 68 76 Pocked Goniometer PGX is ready to use Double click on the device to start the Test software Note If on...

Страница 12: ...the PGX instrument on top of the calibration plate with its reference mark positioned above the dot as indicated in the picture Push gently on top of the instrument to make sure the instrument rests f...

Страница 13: ...e g tissues textile and film or warping materials e g newsprint cigarette tipping paper should be mounted on top of a flat backing material using a soft sticky adhesive material A Place the PGX instru...

Страница 14: ...the pump sequence to inhibit application of the droplet The liquid system is filled in the following way a Place the instrument on top of a flat surface b Locate the tiny air hole on top of the white...

Страница 15: ...ize is adjustable to not interfere with other windows Double click on the image to switch the live image between greyscale and binary black white mode Click with the right button on the image to open...

Страница 16: ...it It allows to close the Camera window even if the software doesn t reply 7 2 Graphmaster test This manual covers only the functions specific for the 68 76 Pocket Goniometer see the GraphMaster manua...

Страница 17: ...ange is always available even when the test substrate appears too dark or too bright due to variations in light reflectance Double click on the label Gain Offset to restore the default value 24 Exposu...

Страница 18: ...Down keys or pick up the base line handle 1 with the mouse and slide it to the correct position Adjust the position of droplet contour When the automatic contour is set incorrectly keep press Shift a...

Страница 19: ...t Defines the minimum height of a droplet reduce for smooth surfaces Default value 10 Sobel Filter Enable the Sobel filter to improve measurements on rough surface but it can increase the stability of...

Страница 20: ...7 2 5 Calibration 7 3 Graphmaster database 7 3 1 Surface energy The Surface Energy functions is available only for the Static and Dynamic mode The function is enabled only when 2 or 3 records are sele...

Страница 21: ......

Страница 22: ...ond impossible to measure Instead the droplet is gently transferred to the surface from a very short free fall distance 8 2 1 Parameters Auto Method Off Manual test Volume Automatic test with droplet...

Страница 23: ...ing tip must be equipped with a fine capillary tube inserted into the PTFE tubing P N 860309 In addition the distance between the dispensing tip and the test surface must be short enough to make sure...

Страница 24: ......

Страница 25: ......

Страница 26: ...e2 are the two times enter as parameters and the Angle1 and Angle2 are the angle measures at the prefix times The end test will be the values of the second time 8 6 1 Parameters Auto Method Off Manual...

Страница 27: ...with the manual results superimposed Use the slider below the image to shift to another image 8 8 Surface free energy SFE As this surface property cannot be measured directly it is calculated from co...

Страница 28: ...let applied on a reflecting background will appear as a dark saucer when the contact angle is below 90 degrees A high contact angle above 90 degrees will appear as a black snowman with a waist against...

Страница 29: ......

Страница 30: ...he pump lid screw C is firmly tightened c If still no liquid is pumped release stop screw B a turn and then tighten the pump lid screw C firmly again d Repeat c above until liquid is pumped inside tub...

Страница 31: ...d operating systems as these are completely outside our control Install a licensed Microsoft version on your computer re install the PG software and the problem should be gone Jumping image With a lim...

Страница 32: ...in contact angle testing for historical reasons In general the contact angle is not affected by the droplet size as long as the droplet is not affected by the gravity For water having a high surface...

Страница 33: ...Adjustment of Pump Volume The micro pump delivers droplets in steps of 0 5 l and for a target volume of 4 0 l the pump will make eight pump movements The Static Mode can be used to measure the average...

Страница 34: ......

Страница 35: ...35...

Страница 36: ...36...

Отзывы: