FIBRO 68-76-00-0001 Скачать руководство пользователя страница 32

|   32

11.1 

Instrument storage

The liquid system (dispenser, tubing and liquid container) should never be 
left with a liquid for a long time as this may clog the system. The procedures 
below are recommended when your PGX+ is to be stored for some time or 
shipped to another location.

When using the PGX+ with liquids other than water (e.g. solvents, inks) it is 
necessary to clean the liquid system after the test has been finished. If the 
instrument will be stored for some time it is also recommended to empty the 
liquid system in the following way:

a) 

Release the pump lid screw to access the pump tubing

b) 

Lift out the liquid container, remove the plug and the empty  

 container
c) 

Blow liquid container dry with air

d) 

Connect syringe with the RED tip to the dispensing tip and pump  

 

air through the tubing until dry

e) 

Push the plug back into liquid container top and re-install the  

 container
f ) 

Put the pump tubing in position

g) 

Re-install the pump lid and secure the pump lid screw gently

NOTE: 
DO NOT tighten the pump lid screw completely when stored!

11.2 

Transportation

Before shipping, the instrument should be prepared as described in above 
section.
Please note it is important to empty even a water-based system completely if 
shipping might include airfreight at high altitude as this might cause freezing 
damages.

Always ship the instrument in a complete case with all instrument 
accessories.

Make sure the case is shipped in sufficient padding to avoid damaging the 
instrument case and its contents.

11  MISCELLANEOUS

 

11.3 

Droplets

11.3.1  Recommended droplet sizes

The default droplet size is 4,0 µl, which has been used as the recommended 
size in contact angle testing for historical reasons. In general the contact 
angle is not affected by the droplet size as long as the droplet is not affected 
by the gravity. For water (having a high surface tension) the contact angle 
will be stable as long as the full droplet remains in view.

The droplet size may still have an input on the reading, as a bigger droplet 
will cover a larger test area and might easier pick up contaminants on an 
inhomogeneous test surface.

For absorbent surfaces it might be easier to track the penetration of liquid 
into the substrate if a slightly bigger droplet is used for the test. On the other 
hand, a bigger droplet may spread outside the field of view or cause the test 
surface to warp leading to erroneous test results.

If the specimen warps when the liquid droplet is applied, it might be 
necessary to attach the specimen strip to a flat surface with a suitable 
adhesive.

11.3.2  Detection of the droplet in Dynamic Mode

The distance between the pendant droplet and the specimen surface is 
necessary for the droplet to form and release from the dispensing tip before 
the droplet touches the surface.

The impact on the contact angle from a falling droplet of 4.0 µl is very small.
This can be verified if a solid, non-absorbent surface is tested for the static 
contact angle with 6-8 droplets. When the same specimen is tested with the 
dynamic approach the contact angles should match the average result from 
the static test.

Smaller droplets will require a bigger falling distance to release from the 
dispensing tip.
A droplet of 2.0 µl falling from a two-mm distance will still give the same 
contact angle reading.

Bigger droplets will give lower contact angles for bigger falling distances.
This principle can be used to determine the “receding “ contact angle 
describing the de-wetting properties of a surface.
 

Содержание 68-76-00-0001

Страница 1: ......

Страница 2: ......

Страница 3: ...ng the eventual transport protections Before shipping an authorization to return should be secured from Fibro System by telephone FAX or written letter The best way to ship will also be advised at tha...

Страница 4: ...4...

Страница 5: ...ial specimens 13 6 3 Pump dispenser 14 7 Software functions 15 7 1 Graphmaster camera 15 7 2 Graphmaster test 16 7 3 Graphmaster database 20 8 Test methods 21 8 1 Static Mode 21 8 2 Dynamic mode 22 8...

Страница 6: ...6...

Страница 7: ......

Страница 8: ...ing the instrument is connected to the special PC software The software gives you an easy overview of the possible tests During the test the data transferred to the computer The pictures taken and the...

Страница 9: ...sta Windows 7 Windows 8 Windows 8 1 or Windows 10 3 3 Scope of supply Your PGX instrument is delivered in a carrying case see picture including PGX software one millilitre syringe calibration plate gl...

Страница 10: ......

Страница 11: ...and I follow the displayed instructions When the installation is terminated the new device 68 76 Pocked Goniometer PGX is ready to use Double click on the device to start the Test software Note If on...

Страница 12: ...the PGX instrument on top of the calibration plate with its reference mark positioned above the dot as indicated in the picture Push gently on top of the instrument to make sure the instrument rests f...

Страница 13: ...e g tissues textile and film or warping materials e g newsprint cigarette tipping paper should be mounted on top of a flat backing material using a soft sticky adhesive material A Place the PGX instru...

Страница 14: ...the pump sequence to inhibit application of the droplet The liquid system is filled in the following way a Place the instrument on top of a flat surface b Locate the tiny air hole on top of the white...

Страница 15: ...ize is adjustable to not interfere with other windows Double click on the image to switch the live image between greyscale and binary black white mode Click with the right button on the image to open...

Страница 16: ...it It allows to close the Camera window even if the software doesn t reply 7 2 Graphmaster test This manual covers only the functions specific for the 68 76 Pocket Goniometer see the GraphMaster manua...

Страница 17: ...ange is always available even when the test substrate appears too dark or too bright due to variations in light reflectance Double click on the label Gain Offset to restore the default value 24 Exposu...

Страница 18: ...Down keys or pick up the base line handle 1 with the mouse and slide it to the correct position Adjust the position of droplet contour When the automatic contour is set incorrectly keep press Shift a...

Страница 19: ...t Defines the minimum height of a droplet reduce for smooth surfaces Default value 10 Sobel Filter Enable the Sobel filter to improve measurements on rough surface but it can increase the stability of...

Страница 20: ...7 2 5 Calibration 7 3 Graphmaster database 7 3 1 Surface energy The Surface Energy functions is available only for the Static and Dynamic mode The function is enabled only when 2 or 3 records are sele...

Страница 21: ......

Страница 22: ...ond impossible to measure Instead the droplet is gently transferred to the surface from a very short free fall distance 8 2 1 Parameters Auto Method Off Manual test Volume Automatic test with droplet...

Страница 23: ...ing tip must be equipped with a fine capillary tube inserted into the PTFE tubing P N 860309 In addition the distance between the dispensing tip and the test surface must be short enough to make sure...

Страница 24: ......

Страница 25: ......

Страница 26: ...e2 are the two times enter as parameters and the Angle1 and Angle2 are the angle measures at the prefix times The end test will be the values of the second time 8 6 1 Parameters Auto Method Off Manual...

Страница 27: ...with the manual results superimposed Use the slider below the image to shift to another image 8 8 Surface free energy SFE As this surface property cannot be measured directly it is calculated from co...

Страница 28: ...let applied on a reflecting background will appear as a dark saucer when the contact angle is below 90 degrees A high contact angle above 90 degrees will appear as a black snowman with a waist against...

Страница 29: ......

Страница 30: ...he pump lid screw C is firmly tightened c If still no liquid is pumped release stop screw B a turn and then tighten the pump lid screw C firmly again d Repeat c above until liquid is pumped inside tub...

Страница 31: ...d operating systems as these are completely outside our control Install a licensed Microsoft version on your computer re install the PG software and the problem should be gone Jumping image With a lim...

Страница 32: ...in contact angle testing for historical reasons In general the contact angle is not affected by the droplet size as long as the droplet is not affected by the gravity For water having a high surface...

Страница 33: ...Adjustment of Pump Volume The micro pump delivers droplets in steps of 0 5 l and for a target volume of 4 0 l the pump will make eight pump movements The Static Mode can be used to measure the average...

Страница 34: ......

Страница 35: ...35...

Страница 36: ...36...

Отзывы: