7.2.3 Elaboration image
The windows is open with a double click on the image from the main page
Elaboration Image allows to change the points where the measurement are
done.
Adjust the position of an incorrect Base line
When the automatic base line is set incorrectly, keep press Shift and use the
“Page Up” and “Page Down“ keys or pick up the base line handle (1) with the
mouse and slide it to the correct position.
Adjust the position of droplet contour
When the automatic contour is set incorrectly, keep press Shift and use the
cursor to move the contour to the correct position.
Adjust the dimension of droplet contour
When the automatic contour is incorrectly, keep press Shift and use the “+”
and “-“keys to change the dimension of the contour to the correct dimension.
These images will then be recalculated with the new base line setting.