XT Extrication Device
TEXTILE COMPONENTS (FLOATS, COVERS, RESTRAINTS)
Make sure that all textile parts and stitching on the device
are free from cuts, abrasion, frayed borders, wear, corrosion,
burns and signs of chemical substances; also check that the
stitching is not cut, frayed or loose; finger textile devices
all along their length to check their integrity. Ensure the
integrity of the stitching protective cover (no movements
or detachments from its terminal). Before use, also check
that the maintenance sheets of each device are correctly
updated. During every use regularly check that the product
works properly and that any other component of the
system is well connected and arranged
; pay attention to
the effects of humidity and frost, as they can compromise
the device operation; avoid the contact with sharp edges
and, if necessary, use protective equipment. As far as the
textile material is concerned, the MAXIMUM useful life is
considered to be equal to 10 years from its first use, so it
must be replaced after this period of time.
8.6 Disinfection
To be carried out after each use, disinfect the surfaces of the
device with a non-abrasive product.
1. Remove restraints, handles, headrest and any other
accessory from the product.
2. In case of surfaces visibly contaminated by organic
substances proceed with decontamination according to the
following procedure:
Remove the material using a disposable cloth and/or paper
and dispose of it in the specific container for special waste
and proceed with cleaning operations.
8.4 Cleaning
The device has to be cleaned as needed
1. Remove any restraint, handle, headrest and in case
accessories.
2. Wash all components of the product with a cloth or
sponge dampened with detergent solution (rinsing the
sponge frequently with clean water).
3. If necessary, remove stains with a stiff bristle brush (not
metal), or a light solvent, paying attention to the labels.
4. Use a cloth to rinse and dry the treated surfaces, make
sure that all the device's components are completely dry
before using them.
5. After cleaning, be sure to refit the device with restraints,
handles, headrest and any other accessory that has been
removed.
Warning: during disinfection, use gloves and any appropriate
PPE. Also ensure there are no signs of wear and/or damage.
8.5 Inspection
To ensure optimal use of the device, it is important to keep all
the components in good conditions and to remove all traces
of dirt, debris and body fluids. The extrication device must
be inspected and cleaned before and after each use. When
not in use, it must be stored in a dry, indoor environment and
inspected at least once a month.
The extrication device should be inspected regularly by
maintenance personnel. If signs of wear or damage are
found during inspection, discontinue use of the device and
immediately contact Ferno's Technical Support Service
to request repairs or maintenance. See section Technical
Support.