![Ferno 24 MAXX Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/ferno/24-maxx/24-maxx_user-manual_554363025.webp)
25
© Ferno-Washington, Inc 234-3307-04 May 2011
24•MAXX®, 24•miniMAXX®, 24H•miniMAXX®
Maintenance
7 - MainTenanCe
7.4 Disinfecting and Cleaning
the Cot
To disinfect:
Remove all restraints and the mattress from
the cot. Wipe all surfaces with disinfectant following the
disinfectant manufacturer’s directions for application
method and contact time. Ferno recommends that you
inspect the cot for obvious damage as you disinfect it.
To clean:
Hand clean all surfaces of the cot with warm
water and a mild detergent. Rinse with warm, clear water.
Dry with a towel or allow the cot to air-dry.
7.1 Maintenance schedule
The cot requires regular maintenance. Set up and follow
a maintenance schedule. The table at right represents
minimum intervals for maintenance.
Keep maintenance records. A sample maintenance record
sheet is provided on page 31.
When using maintenance products, follow the manufacturers’
directions and read the manufacturers’ material safety data
sheets. You can purchase a recommended disinfectant from
your Ferno distributor or Ferno Customer Relations (page 2).
7.2 Disinfecting and Cleaning
the restraints
Remove the restraints from the cot. Disinfect and clean only
as directed in the restraint users’ manuals provided with the
restraints. Additional, free users’ manuals can be obtained
from Ferno Customer Relations (page 2).
Minimum
Maintenance intervals
each U
se
a
s
n
eeded
each M
on
th
Disinfecting
(this page)
•
Cleaning
(this page)
•
Waxing
(page 26)
•
inspecting
(page 26)
•
•
lubricating
(page 27)
•
7.3 Disinfecting and Cleaning
the Mattress
1. Remove the mattress from the cot.
2.
To disinfect:
Apply disinfectant to the mattress
according to the disinfectant manufacturer’s
instructions for application method and contact time.
3.
To clean:
Wash the mattress with warm, soapy water
and a soft cloth. Rinse the mattress with clear water.
Hang the mattress to dry, or dry it with a towel.
important
Disinfectants and cleaners containing bleach,
phenolics, or iodines can cause damage. Do not use
products containing these chemicals.
Warning
improper maintenance can cause injury. Maintain
the cot only as described in this manual.
important
Using abrasive cleaning compounds or applicators
on the cot can cause damage. Do not use abrasive
materials to clean the cot.
important
Water under high pressure, or steam, can penetrate
joints, flush away lubricant, and cause corrosion. Use
caution when cleaning moving parts such as joints and
hinges, and reapply lubricant if needed (see page 27).