Note: 

 

Values written to the above holding registers remain after Power-off/Reset 

 

To perform a span calibration for 

CO2

C2H4

 or 

O2

 sensor: 

o

 

Apply 

C

alibration gas

 at device inlet 

o

 

Wait until reading stabilized  

o

 

Read sensor’s current measurement on 

Input Registers

 

and sensor’s SPAN value on 

Holding Registers

 

o

 

Calculate Span value using the provided 

formula

 and round this value to the nearest integer 

o

 

Write new span value back to the sensor’s SPA

N register 

 

To perform a zero calibration for 

C2H4

 or 

O2

 sensor (see also zero calibration with coils setting bellow): 

o

 

Put device in Continuous measurement mode (write 0 to TRIGGER_INTERVAL register) 

o

 

Apply 

Zero gas

 at device inlet  

o

 

Wait until reading stabilized (C2H4 >= 7 minutes, O2 >= 2 minutes) 

o

 

Read sensor RAW value o

Input Registers

 

o

 

Write this value to the according C2H4/O2 ZERO register 

o

 

Put device back in Trigger measurement mode (write 300 to TRIGGER_INTERVAL register) 

Coils 

Mode: Read/Write, size: 1 bit 

Name 

Address 

Default  Description 

ZERO_C2H4 

False 

Request to zero C2H4 sensor 

ZERO_O2 

False 

Request to zero O2 

ZERO_CO2 

False 

Request to zero CO2 sensor 

ZERO_CONF 

False 

Confirmation of zero action 

FLOW_MODE 

True 

Set the air flow configuration. 
True: flow in sampling mode 
False: flow in cleaning mode 

RESET 

False 

True: (software) reset  
False: no action 

 

IMPORTANT

:

 backup calibration parameters before performing a zero calibration. A zero calibration will overwrite 

the factory zero calibration parameter in Holding register.  

Note:  

 

To perform a zero calibration using coils for 

CO2

C2H4

 or 

O2

 sensor 

o

 

Put device in Continuous measurement mode (write 0 to TRIGGER_INTERVAL register) 

o

 

Apply 

Zero

 gas

 at device inlet 

o

 

Wait until reading stabilized (C2H4 >= 7 minutes, CO2/O2 >= 2 minutes) 

o

 

Set ZERO_[

SENSOR

] coil (ZERO_C2H4/ZERO_O2/ZERO_CO2) to request zero action 

o

 

Set  ZERO_CONF  coil  to  confirm.  After  confirmation,  the  device  will  perform  zero  action  with  the 
current gas and automatically reset the ZERO_[

SENSOR

] and ZERO_CONF coil. New zero values will 

also be updated on Holding Registers 

o

 

Put device back in Trigger measurement mode (write 300 to TRIGGER_INTERVAL register) 

 

 

Содержание F-901B

Страница 1: ...July 2019 F 901B Reference Manual Version 1 4 6 15 16 1554 NE 3rd Ave Camas WA 98607 USA Phone 360 833 8835 www felixinstruments com...

Страница 2: ...6 Operating modes 6 MODBUS 7 RTU RS485 Configuration 7 Pin Description 7 Absolute Maximum Rating 8 RS485 Modbus Parameters 8 F 901B MODBUS Specifications 8 Input Registers 8 Holding Registers 9 Coils...

Страница 3: ...e socket If sampling via tubing make sure that the tubes are securely attached to the device before operating Use the included hydrophobic filter to prevent liquid water from entering device See Cable...

Страница 4: ...m Lower Detection Limit 2 ppm Response Time T90 3 minutes Temperature Range 0 C to 50 C Pressure Range 1013mbar 10 Relative Humidity Range 15 to 90 R H non condensing Long Term Output Drift 5 per mont...

Страница 5: ...CURED BEFORE POWERING UP DEVICE MAKE SURE THERE IS NO FLOW OBSTRUCTION AT INLET OUTLET BEFORE POWERING UP DEVICE USE HYDROPHOBIC FILTER DURING SAMPLING TO PROTECT THE SENSOR FROM CONDENSATION Waterpro...

Страница 6: ...ts not supplied Part number Description Manufacture 767KS12 DC Power Plug Sealed IP68 Switchcraft 630125673867 Patch Cable Cap IP67 water and dust protection InstallerParts Lock Ring 12VDC GND Plug ha...

Страница 7: ...cycle for 30 seconds The whole routine will then repeat after some idle time Final sensor readings are updated at the end of sampling cycle and remain the same until the next update see TRIGGER_INTEV...

Страница 8: ...n Pin Name Type Description 1 3 3V Reference potential 3 3V Reference Voltage 2 GND Reference potential Local device ground 3 UART_TX Digital Output Firmware update interface 4 B D Bus In Out Driver o...

Страница 9: ...te Single Register o 0x0F Write Multiple Coils o 0x10 Write Multiple Registers Exception messages supported Default address o RTU 50 o TCP IP 192 168 1 50 Input Registers Mode Read only size 16 bits N...

Страница 10: ...ibration parameter TRIGGER_INTERVAL 20 0 Interval between Trigger measurements in seconds Writing a value greater than 0 to this register enables Trigger mode see Operating modes Writing a 0 to this r...

Страница 11: ...rite size 1 bit Name Address Default Description ZERO_C2H4 4 False Request to zero C2H4 sensor ZERO_O2 5 False Request to zero O2 ZERO_CO2 6 False Request to zero CO2 sensor ZERO_CONF 7 False Confirma...

Страница 12: ...ic discharge protection Follow the below steps carefully to ensure proper sensor installations Step 2 Apply even forces at both ends to pull the Filter Tube off connectors Step 4 Pull C2H4 sensor off...

Страница 13: ...ove old C2H4 sensor from connector Plug in the new C2H4 sensor then push into socket push into socket Step 7 screw new O2 sensor into socket then plug in connector Step 8 Apply force to both ends gent...

Страница 14: ...on typically involves a zero baseline and a span sensitivity calibration Zero calibration commonly uses zero gas such as 100 N2 while span calibration uses target gas for calibration Below are recomme...

Страница 15: ...re to be covered only by the warranty of its manufacturer Felix Instruments will not be liable to the Buyer for loss damage or injury to persons or to property by the use of equipment manufactured by...