background image

F-901B Reference Manual 

Page 

9

 of 

15

 

 

DEV_TYPE 

100 

Default device type ID: 9011 

FIRMWARE 

101 

Firmware version 

 

Holding Registers 

Mode: Read/Write, size: 16 bits 

Name 

Address 

Default 

Description 

PUMP_POWER 

50 

Internal pump power 0-100% 

C2H4_SPAN 

 

C2H4 span calibration parameter, calculated as follow: 
C2H4_SPAN = C2H4_CUR*C2H4_SPAN_CUR/C2H4_CAL 
Note: C2H4_CAL: expected calibration concentration 
           C2H4_SPAN_CUR: Current span value 
           C2H4_CUR: Current C2H4 measurement 

C2H4_ZERO 

 

C2H4 zero calibration parameter 

O2_SPAN 

 

O2 span calibration parameter, calculated as follow: 
O2_SPAN = O2_CUR*O2_SPAN_CUR/O2_CAL 
Note: O2_CAL: expected calibration concentration 
           O2_SPAN_CUR: Current span value 
           O2_CUR: Current O2 measurement 

O2_ZERO 

 

O2 zero calibration parameter 

CO2_SPAN 

 

CO2 span calibration parameter, calculated as follow: 
CO2_SPAN = CO2_CAL*CO2_SPAN_CUR/CO2_CUR 
Note: CO2_CAL: expected calibration concentration 
           CO2_SPAN_CUR: Current span value 
           CO2_CUR: Current CO2 measurement 

CO2_ZERO 

 

CO2 zero calibration parameter 

TRIGGER_INTERVAL 

20 

Interval between Trigger measurements in seconds.  
Writing a value greater than 0 to this register enables 
Trigger mode (se

Operating modes

). Writing a 0 to this 

register disables the Trigger mode (switches to 
Continuous mode after finishing any on-going Trigger 
measurement). Note that a single Trigger mode 
measurement takes at least 65 seconds. 

SLAVE_ADDR 

30 

50 

Device Modbus RTU slave address. Update this register 
to change the slave address. Valid slave addresses: 0-99 

MODBUS_MODE 

31 

0: TCP/IP (LED blinks 2 times at boot) 
1: RTU (LED blinks 3 times at boot) 
2: Auto Config Mode. During boot up, if the internal 
TCP/IP cable is plugged in and the device is connected to 
active ethernet hub/router or the F901C Controller, the 
device will select TCP/IP mode. Otherwise, RTU mode is 
selected. 

IP_ADDR0 

32 

192 

Device IP address 

IP_ADDR1 

33 

168 

Device IP address 

IP_ADDR2 

34 

Device IP address 

IP_ADDR3 

35 

50 

Device IP address 

 

IMPORTANT

:

 backup calibration parameters before overwriting their values (performing a calibration) or update 

device firmware (all parameters will be erased). All F-901B comes with factory calibration using standard certified 
gases.  

Содержание F-901B

Страница 1: ...July 2019 F 901B Reference Manual Version 1 4 6 15 16 1554 NE 3rd Ave Camas WA 98607 USA Phone 360 833 8835 www felixinstruments com...

Страница 2: ...6 Operating modes 6 MODBUS 7 RTU RS485 Configuration 7 Pin Description 7 Absolute Maximum Rating 8 RS485 Modbus Parameters 8 F 901B MODBUS Specifications 8 Input Registers 8 Holding Registers 9 Coils...

Страница 3: ...e socket If sampling via tubing make sure that the tubes are securely attached to the device before operating Use the included hydrophobic filter to prevent liquid water from entering device See Cable...

Страница 4: ...m Lower Detection Limit 2 ppm Response Time T90 3 minutes Temperature Range 0 C to 50 C Pressure Range 1013mbar 10 Relative Humidity Range 15 to 90 R H non condensing Long Term Output Drift 5 per mont...

Страница 5: ...CURED BEFORE POWERING UP DEVICE MAKE SURE THERE IS NO FLOW OBSTRUCTION AT INLET OUTLET BEFORE POWERING UP DEVICE USE HYDROPHOBIC FILTER DURING SAMPLING TO PROTECT THE SENSOR FROM CONDENSATION Waterpro...

Страница 6: ...ts not supplied Part number Description Manufacture 767KS12 DC Power Plug Sealed IP68 Switchcraft 630125673867 Patch Cable Cap IP67 water and dust protection InstallerParts Lock Ring 12VDC GND Plug ha...

Страница 7: ...cycle for 30 seconds The whole routine will then repeat after some idle time Final sensor readings are updated at the end of sampling cycle and remain the same until the next update see TRIGGER_INTEV...

Страница 8: ...n Pin Name Type Description 1 3 3V Reference potential 3 3V Reference Voltage 2 GND Reference potential Local device ground 3 UART_TX Digital Output Firmware update interface 4 B D Bus In Out Driver o...

Страница 9: ...te Single Register o 0x0F Write Multiple Coils o 0x10 Write Multiple Registers Exception messages supported Default address o RTU 50 o TCP IP 192 168 1 50 Input Registers Mode Read only size 16 bits N...

Страница 10: ...ibration parameter TRIGGER_INTERVAL 20 0 Interval between Trigger measurements in seconds Writing a value greater than 0 to this register enables Trigger mode see Operating modes Writing a 0 to this r...

Страница 11: ...rite size 1 bit Name Address Default Description ZERO_C2H4 4 False Request to zero C2H4 sensor ZERO_O2 5 False Request to zero O2 ZERO_CO2 6 False Request to zero CO2 sensor ZERO_CONF 7 False Confirma...

Страница 12: ...ic discharge protection Follow the below steps carefully to ensure proper sensor installations Step 2 Apply even forces at both ends to pull the Filter Tube off connectors Step 4 Pull C2H4 sensor off...

Страница 13: ...ove old C2H4 sensor from connector Plug in the new C2H4 sensor then push into socket push into socket Step 7 screw new O2 sensor into socket then plug in connector Step 8 Apply force to both ends gent...

Страница 14: ...on typically involves a zero baseline and a span sensitivity calibration Zero calibration commonly uses zero gas such as 100 N2 while span calibration uses target gas for calibration Below are recomme...

Страница 15: ...re to be covered only by the warranty of its manufacturer Felix Instruments will not be liable to the Buyer for loss damage or injury to persons or to property by the use of equipment manufactured by...

Страница 16: ...ax _____________________________ Felix Instruments Serial Number s ________________________________________ Purchase Date _____________________Purchase Price _______________________ Your opinions will...

Отзывы: