filter can help reduce the interference measured when alcohol or other light polar molecules
(known interfering gases for the F-900 electrochemical sensor) are present. Also, using the
water trap can help keep the electrochemical cell hydrated, with longer periods of time between
needing to refill the electrochemical sensor.
There are 3 options for Chamber In in the Setup>Chamber menu.
1. Chamber In set to Off.
2. Chamber In set to On.
a. To use soda lime or silica gel to condition the gas stream before it is
measured, turn Chamber In to On and place the consumable in the
conditioning chamber.
3. Chamber In set to
Special
.
Use the up/down arrows to select Chamber Out or Chamber In. Use the right arrow to get to
Off/On. Next, use the up/down arrow to switch between On or Off, or set the Measure or
Cleaning period in minutes. When Chamber In is set to Special, you will see Measure and
Cleaning Period appear below.
Fill Chamber In with distilled water to max fill line (located on back of F-900)
This mode may not be applicable at high ethylene concentrations.
When the F-900 is using the water to filter out alcohol, the water must be cleaned by
running a longer cleaning period than the measure period.
To use the PolarCept water trap, set Chamber In to
Special
. Fill Chamber In with approximately
35mL of DI water. Do not fill beyond the max fill line on the back of the instrument. Next, set the
length of the measuring period in minutes. This is the time the F-900 will take to do a
measurement. When using the water trap, a stabilizing time of at least 3 minutes is required
before being able to record a measurement. However, a longer period of measurement (6-7
minutes) is required for accurate readings beyond the stabilization period. Actual times will
depend on individual application/environment.
The length of time required for cleaning will be dependent on the make-up of the gas(es) being
sampled. The more interfering gases that are present, the longer the water will need to be
cleaned. The main reason for the cleaning step is to keep the water in the maximum adsorption
Chamber Menu
CHAMBER_A OUT Off
CHAMBER_B IN SPECIAL
MEASURE PERIOD 6 MIN
CLEANING PERIOD 17 MIN