Using   the   Falcon   II   on   the   Tecnai   F20   

   

1.

 

Perform   all   microscope   alignments   out   of   Low   Dose   

2.

 

Make   sure   TIA   is   open   

3.

 

Activate   Low   Dose   and   load   the   50K_recommended.lds   file   or   set   up   
your   low   dose   as   below:   

   

   

   
At   50,000x   the   pixel   size   is   2.08Å

 

Search 

Focus 

Exposure (2 sec 

integration time) 

Magnification 

(Detector) 

1,700x (Orius) 

 

150,000x (Orius) 

 

50,000x (Falcon) 

C2 aperture 

100 

100 

100 

Spot size 

Defocus 

-200 

µ

0-4 

µ

 

Dose Rate 

(Measured exposure) 
on small screen)

 

<0.02 e-/Å

2

/sec 

 

10 e-/Å

2

/sec (2.7 sec) 

Содержание Tecnai G2 F20

Страница 1: ...the ones below and highlighted in green Gun will always be 1 Log 4 Fill the cold trap dewar with LN2 Cover the screen window first Fill the dewar until it is full If you are the first user of the day...

Страница 2: ...ding Remove the objective aperture by sliding the lever to the right Under Setup tab click on Turbo to switch on the turbo pump The turbo pump will start speeding up orange button color and in 3 minut...

Страница 3: ...ol panel 2 Centering the condenser aperture Condenser a Select a C2 aperture 100um for TEM Single particle b At a magnification of 10 000 go to crossover and center the beam using the left trackball c...

Страница 4: ...e phosphor screen using Multifunction knobs X and Y b Beam Tilt Pivot Points X and Y Click on Beam Tilt Pivot Points X and go to crossover Try to minimize the movement of the beam using Multifunction...

Страница 5: ...sert the objective aperture in slide left and using the aperture position knobs on the aperture center it around the center of the un diffracted beam 6 Objective lens stigmation To stigmate the object...

Страница 6: ...k Search or Preview to view the image and click Focus and move the red box to a contrasty area to adjust focus using the objective focus knob i Click Acquire to obtain an image In the TIA analysis mod...

Страница 7: ...the 50K_recommended lds file or set up your low dose as below At 50 000x the pixel size is 2 08 Search Focus Exposure 2 sec integration time Magnification Detector 1 700x Orius 150 000x Orius 50 000x...

Страница 8: ...Appendix Microscope Information Line resolution nm 0 144 Cs objective mm 2 0 Cc objective mm 2 0 Focal length mm 2 7 Detector FEI Falcon II Physical pixel size m 14...

Отзывы: