Rotation Center alignment: 

Find an identifiable object in your specimen. 

Click on 

Rotation Center 

and minimize its displacement using MF X and Y.  

(d) 

Comma free alignment:

 Use the Orius CCD to go slightly underfocus (-500 

nm). Using the live FFT correct 

Direct Alignments

Comma free alignment X 

and Y

. Make sure the alternating power spectra show the zeros at the same 

frequencies. 

 

Once you have performed the Direct Alignments click on 

Done 

to exit. 

   

 

5. Centering the objective aperture.  

At a magnification of 50000x press the diffraction button on the 

right control panel. Insert the 

objective aperture in (slide left) and using the aperture position knobs on the aperture 
center it around the center of the un-diffracted beam.

 

   

   

6.  Objective lens stigmation. 

   

To stigmate the objective lens go to 50,000X, start the Orius CCD and set the focus to 
slightly underfocus on the carbon support and bring up a live FFT so as you can see 2-3 
Thon rings. If astigmatism is present these rings will be oval. To make the rings circular go 
to the 

Tune

 tab in the 

Stigmator

 panel, click on 

Objective

 

to activate the objective 

stigmators. Use MF X and Y knobs, one at a time, to make the rings circular. If the rings 
are not visible due to severe astigmatism either reduce the magnification or defocus further 
and then stigmate as before. Once finished click on 

None

Содержание Tecnai G2 F20

Страница 1: ...the ones below and highlighted in green Gun will always be 1 Log 4 Fill the cold trap dewar with LN2 Cover the screen window first Fill the dewar until it is full If you are the first user of the day...

Страница 2: ...ding Remove the objective aperture by sliding the lever to the right Under Setup tab click on Turbo to switch on the turbo pump The turbo pump will start speeding up orange button color and in 3 minut...

Страница 3: ...ol panel 2 Centering the condenser aperture Condenser a Select a C2 aperture 100um for TEM Single particle b At a magnification of 10 000 go to crossover and center the beam using the left trackball c...

Страница 4: ...e phosphor screen using Multifunction knobs X and Y b Beam Tilt Pivot Points X and Y Click on Beam Tilt Pivot Points X and go to crossover Try to minimize the movement of the beam using Multifunction...

Страница 5: ...sert the objective aperture in slide left and using the aperture position knobs on the aperture center it around the center of the un diffracted beam 6 Objective lens stigmation To stigmate the object...

Страница 6: ...k Search or Preview to view the image and click Focus and move the red box to a contrasty area to adjust focus using the objective focus knob i Click Acquire to obtain an image In the TIA analysis mod...

Страница 7: ...the 50K_recommended lds file or set up your low dose as below At 50 000x the pixel size is 2 08 Search Focus Exposure 2 sec integration time Magnification Detector 1 700x Orius 150 000x Orius 50 000x...

Страница 8: ...Appendix Microscope Information Line resolution nm 0 144 Cs objective mm 2 0 Cc objective mm 2 0 Focal length mm 2 7 Detector FEI Falcon II Physical pixel size m 14...

Отзывы: