4000 Series
Page
7
of
41
3460006-17_serial series 5000000
z.
Drain: External with heated evaporator pan, powered from the cabinet.
aa.
Water Supply for water cooled condenser model 4102/4105
Tower water at ≤ 85F (29.5C) and ≥ 20 psi differential (1.38 Bar):
Water flow rate at maximum load = 7.0 gpm (26.5 liters/ min)
Water flow rate at average load = 4.5 gpm (15.0 liters/ min)
Chilled water at ≤ 50F (10C) and ≥ 20 psi differential (1.38 Bar):
Water flow rate at maximum load = 3.5 gpm (13.3 liters/ min)
Water flow rate at average load = 2.0 gpm (7.6 liters/ min)