4000 Series
Page
24
of
41
3460006-17_serial series 5000000
Wait For Process 1
-80ºC
Wait Event 1
Off
Wait Event 2
Off
Day of Week
Every Day
Jump Step
1
Jump Count
1
End Type
User
Event 1
On
Tied to internal outputs on the controller
Event 2
Off
Event 1 must be ON and Event 2 OFF
for this step type
Soak Step
This step type will hold the cabinet at the set point temperature defined in the previous step, for the specified time.
Step Type
Soak
Target Set Point Loop 1
0ºC
Target Set Point Loop 2
-18ºC
Hours
0
Use these parameters to set the soak
Minutes
10
time
Seconds
0
Rate
0ºC
Wait for Process Instance
1
Wait For Process 1
0ºC
Wait Event 1
Off
Wait Event 2
Off
Day of Week
Every Day
Jump Step
1
Jump Count
1
End Type
User
Event 1
On
Tied to internal outputs on the controller
Event 2
Off
Event 1 must be ON and Event 2 OFF
for this step type
Wait for Process
This step type is used to guarantee the chamber temperature is at a specific temperature before moving on to the
next step in the profile. The “Wait for Process 1” parameter is where this temperature is defined at.
Step Type
Wait For Process
Target Set Point Loop 1
-80ºC
Target Set Point Loop 2
-18ºC
Hours
0
Minutes
0