Water cooled condenser manually test or flushing the valve (Water cooled only
4102/4105)
The water-cooled condenser water regulating valve should be inspected every six (6)
months. This is a visual and operational inspection, complete the following steps:
1. Turn off power to the freezer.
2. Remove side panel cover (20 screws)
3. Manually flush the valve and piping before and after installing, repairing, or replacing
a valve to remove filings, chips, or other foreign matter. Manual flushing does not
affect valve adjustment. To flush the valve, insert screwdrivers under both sides of
the valve spring guide and lift upwards. (see figure 1 below)