54
AUT
OCLA
VE SERIE SK07
MAINTENANCE
Before proceeding with the operations indicated below, always switch off the
main switch (1 Fig. 1) and check the chamber cools.
Use is forbidden of hydrochloric acid and detergents containing chlorine to clean any part
of the autoclave.
- Cleaning the sterilisation chamber and the accessories inside
Check:
Monthly or every 50 cycles.
Use a soft cloth soaked in water so that the protective surface treatment of the chamber is
not damaged. Rinse with plenty of water and, at the end of the operation, dry. If necessary,
dampen the cloth with white vinegar to remove any limescale.
You are however advised to clean the chamber only if strictly necessary.
If limescale deposits are present, immediately check what is causing them. The formation of
white patches at the bottom of the chamber means poor quality water is being used or water
with incorrect characteristics compared to those required.
- Porthole cleaning
Check:
Daily or every 5 cycles.
Use a soft cloth soaked in water to avoid damaging the protective treatment of the porthole.
MAINTENANCE AND CONTROL PLAN
SMALL STERILIZER MODEL SK 07
MAINTENANCE OPERATIONS
Frequency
Operation
Daily
5 cycles
Monthly
50 cycles
Quarterly
500 cycles
Yearly
1000 cycles
Cleaning of hatch
X
Cleaning of hatch sealing
X
Cleaning of camera and trays
X
Cleaning of front door
X
Cleaning of the filter of the camera
X
Adjustment of hatch
Replacement of the batteriological filter
X
Replacement of the hatch gasket
X
PERIODICAL CONTROLS
Frequency
Operation
Before first
Use
Daily
Yearly
2 Years
Vacuum test EN 17665
X
Bowie & Dick Test EN 17665
X
Helix Test EN 17665
X
Safety Check (electrical) - EN 62353
-
Protective Hearth continuity
-
Insulation test
-
Enclosure leakage current
X
Initial Validation EN 17665
X
Periodic Validation EN 17665
X
Check of the safety valve
X
MAINTENANCE AND CONTROL PLAN
SMALL STERILIZER MODEL SK 07
MAINTENANCE OPERATIONS
Frequency
Operation
Daily
5 cycles
Monthly
50 cycles
Quarterly
500 cycles
Yearly
1000 cycles
Cleaning of hatch
X
Cleaning of hatch sealing
X
Cleaning of camera and trays
X
Cleaning of front door
X
Cleaning of the filter of the camera
X
Adjustment of hatch
Replacement of the batteriological filter
X
Replacement of the hatch gasket
X
PERIODICAL CONTROLS
Frequency
Operation
Before first
Use
Daily
Yearly
2 Years
Vacuum test EN 17665
X
Bowie & Dick Test EN 17665
X
Helix Test EN 17665
X
Safety Check (electrical) - EN 62353
-
Protective Hearth continuity
-
Insulation test
-
Enclosure leakage current
X
Initial Validation EN 17665
X
Periodic Validation EN 17665
X
Check of the safety valve
X
Frequency
Frequency
Daily
5 cycles
Before
first use
Monthly
50 cycles
Daily
Quarterly
500 cycles
Annually
Annual
1000 cycles
Every two
years
Porthole cleaning
Vacuum Test EN 17665
Porthole gasket cleaning
Bowie & Dick Test 17665
Chamber and accessory cleaning
Helix Test EN 17665
Front panel cleaning
Front panel cleaning
- Safety earthing continuity
- Insulation test
- Casing dispersion current
Chamber filter cleaning
Initial validation EN 17665
Porthole door adjustment
Periodic validation EN 17665
Bacterial filter replacement
Safety valve check
Porthole gasket replacement
Operations
Operations
PErIoDS CoNTroLS