Press AUTO again to modify the value
to equal the volume of the calibrated
container.
Press key “10” to increase the indicat
-
ed value.
Press key “0.1” to decrease the indicat
-
ed value.
After correction:
Press AUTO again, the meter will calcu
-
late the K Factor automatically which
is blinking.
The meter stores the new Calibration
Factor (K Factor) and displays:
Blinking New K Factor
After a few seconds, the Meter auto
-
matically exits from the customization
mode and starts using the new Calibra
-
tion Factor.
One dispensing operation is enough to
carry out the on-site calibration of the
meter.
If you wish to verify the result of this cal
-
ibration, complete a normal dispensing
in the same Calibrated Container with
-
out re-entering the calibration function
in the Customization menu.
Attention: In order to preserve the unit
setting, OK/STOP must be held for two
seconds after accomplishing setting
and program will be saved until the
start of next function setting. If setting
preservation in flash in case of pow
-
er-off is needed, it is essential to wait for
30 seconds without operation until the
screen disappears and the setting will
written in flash then.
WARNING
After dispensing, wait a
few minutes to allow any
air bubbles in the calibrat-
ed container to dissipate.
Verify the actual amount
dispensed in the calibrated
container; as the air bub
-
bles dissipate, the indicated
level may decrease. Do not
wait more than 15 minutes
as the meter will exit from
the menu and it will no lon-
ger be possible to complete
the calibration operation.
The meter is ready to accept the
modification of the indicated value
to correspond to the real value. A key
each time is pressed, the last digit on
the right is modified under one unit.