![FALEX 400 Скачать руководство пользователя страница 32](http://html.mh-extra.com/html/falex/400/400_operation-and-maintenance-manual_540455032.webp)
31
Section 4: Disassembly and Component Cleaning
4.2 Continued Unit Disassembly
17.
Use the hex wrench to remove the three screws in
the pre-filter assembly.
18.
Disassemble the pre-filter. Remove and discard
the white pre-filter element. Place the remaining
components of the pre-filter assembly into the
provided pan. The pre-filter O-ring is reusable
after proper cleaning.
19.
Open the cover from the sample container area.
20.
Remove the aeration line and the sample-in line
and place them in the pan provided.
21.
Remove the fuel thermocouple from the beaker
and wipe down.
22.
Remove the sample-in and sample-out
glass beakers.
23.
Dispose of fuel appropriately according to
local regulations.
4.3 Component Cleaning
NoTe:
The heater tube housing is the only component
to be cleaned with Tri-Solvent.
1.
Brush the inside of the heater tube housing with
the provided cleaning brush and Tri-Solvent.
2.
Inspect the inner surfaces of the heater tube
housing for cleanliness and repeat step 1
if needed.
3.
Rinse with Iso-Octane and dry using
compressed air.
4.
Use a wash bottle filled with Iso-Octane to flush/
rinse all components in the pan and dry using
compressed air. Make certain all fuel has been
removed from all components.
Содержание 400
Страница 48: ...47 Addendum B Printed Report Example ...
Страница 49: ...48 Notes issue date 1 1 2013 ...