21
22.
Attach the pre-filter assembly to the assembled
test section using the fuel input line
[Figure 25]
.
Inspect the O-rings on both ends of the sample
line union. Replace if missing or damaged.
23.
Finger tighten the fittings until snug.
24.
Plug in the heater tube thermocouple. Carefully
insert the thermocouple tip into the hole on the top
of the heater tube and lower until it stops. Rotate
the thermocouple retaining clip to the left to lock
the thermocouple into place
[Figure 26a and 26b]
.
NoTe:
The heater tube thermocouple can be stored in
the hole found on the top right of the beaker stand to
the left of the bus bar area when not in use.
25.
Check all of the fittings to ensure they are
finger tight.
26.
Close the acrylic cover to the sample test area.
Leaving the cover open will cause an alarm while
a test is running.
If this is the initial setup or if the F400 has not been oper-
ated for 15 days, it is recommended that the automated
Purge be run prior to any tests being performed on the
unit. Refer to section 6.3.3.
Figure 25
Fuel Input Line to Assembled
Test Section
Figure 26a
Thermocouple Retaining Clip
Assembled
Test Section
Fuel Input Line
Thermocouple
Retaining Clip
Retaining
Clip in Place
Pre-Filter Assembly
Insert Thermocouple
into top of Tube
Section 2: Setup
Heater Tube
Thermocouple
Installed
Bus Bar Screws
Figure 26b
Heater Tube Thermocouple and
Retaining Clip