47
EN
12.0 INSTRUCTIONS WITH REGARD TO EMC
•
Medical electrical devices are subject to special precautions with regard to EMC and must be installed in accordance with the
EMC instructions specified below.
•
Portable and mobile HF communication equipment can influence medical electrical devices.
•
The use of accessories, transducers or cables other than those specified can lead to increased emission or a reduced
electromagnetic immunity of the device or system.
12.3 GUIDELINES AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC EMISSIONS
TRACHEOPORT
®
JUNIOR is intended for operation in an environment as that specified below. The customer or user of the
TRACHEOPORT
®
JUNIOR device should ensure that the device is operated in such an environment.
Interference emission measu-
rements
Accordance level
Electromagnetic environment - guideline
HF emissions as per CISPR 11
Group 1
TRACHEOPORT
®
JUNIOR uses HF energy exclusively for its
internal function. Therefore, its HF emission is very low and it is
unlikely that neighbouring electronic equipment will be disrup-
ted.
HF emissions as per CISPR 11
Class B
TRACHEOPORT
®
JUNIOR is suitable for use in all facilities,
including facilities in living quarters and facilities which are
directly connected to a public power supply that also supplies
buildings used for residential purposes
Emission of harmonics as per
IEC 61000-3-2
Not applicable
Emission of voltage fluctuations/
flicker as per IEC 61000-3-3
Not applicable
12.4 RECOMMENDED SAFETY DISTANCES BETWEEN PORTABLE AND MOBILE HF
TELECOMMUNICATION DEVICES AND TRACHEOPORT
®
JUNIOR
Recommended safety distances between portable and mobile HF telecommunication devices
and TRACHEOPORT
®
JUNIOR
TRACHEOPORT
®
JUNIOR is intended for operation in an electromagnetic environment with controlled HF disturbances. The
customer or the user of the TRACHEOPORT
®
JUNIOR can help avoid electromagnetic interference by keeping the minimum
distance between portable and mobile HF telecommunication devices (transmitters) and the
TRACHEOPORT
®
JUNIOR device – depending on the output power of the communication device as specified below.
Safety distance in metres depending on the transmission frequency
Nominal
power of the
transmitter
W
150 kHz to 80 MHz
d = [0.35] √ P
80 MHz to 800 MHz
d = [0.35] √ P
800 MHz to 2.5 GHz
d = [0.70] √ P
0.01
0.035
0.035
0.07
0.1
0.11
0.4
0.22
1
0.35
0.35
0.7
10
1.1
1.1
2.2
100
3.5
3.5
7.0
For transmitters of which the maximum nominal power is not indicated in the table above, the recommended safety distance
d in metres (m) can be determined using the equation that belongs to the respective column, P being the maximum nominal
power of the transmitter in watt (W) as specified by the manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not be applicable in all cases. The propagation of electromagnetic values is influenced
by absorptions and reflections of buildings, objects and people.