10.0 TECHNICAL DATA
Suction capacity of unit
24 ± 2 l/min
max. vacuum at sea level
-76 kPa* (-760 mbar; -570 mmHg) ± 4 kPa
Vacuum indicator
-1...0 bar (± 25 mbar) (mm Hg; kPa)*
Vacuum control
continuously adjustable
-20 kPa* (min.) to -75 kPa* (max.)
Auxiliary air control
manually via Fingertip
Secretion container
600 ml secretion container
Suction tube
ø 6 mm, approx. 1.80 m long
Power supply
(power supply)
100-240 Vac (+/-10%) 50/60 Hz
Low voltage power supply
12 V DC ± 10 %
Operating time with battery
(without mains power supply)
Operating time of battery approx. 60 min
Permanent operation
Operating time with 12V DC
mains power supply (via
power supply or 12V car
supply)
Permanent operation
Emergency operation
If the battery has been completely discharged, the device can operate with mains
power supply
Current consumption
(power supply)
up to 4 A
Power consumption
48 VA
Rechargeable battery
11.1 V; 4.4 Ah; Li ions; at least 500 charging cycles, with partial discharge correspondingly
more; up to 800 - 900 charging cycles in total
Charging time
approx. 4 hrs.
Noise level
55.0 dB (A) @ 1 m
Ambient conditions
Transport/storage
-20...+60°C
10...95 % air humidity without condensation at an air pressure of 700...1060 hPa
Operation
+10...+40°C
30...75 % air humidity without condensation at an air pressure of 700...1060 hPa
Dimensions, HxWxD
370 x 290 x 150 mm
Weight
2.4 kg incl. battery
Protection class (EN 60601-1) II
Applied Part
Type BF
Protection type
IP 21
* 1 bar = 750.06 mm Hg =1000 hPa / depending on prevailing air pressure and ambient conditions.