UM-X
Page 24/35
6.9.1
6.9.1
6.9.1
6.9.1
6.9.1 Settings menu Change
Settings menu Change
Settings menu Change
Settings menu Change
Settings menu Change
The "Change" settings menu displays the set filling level in
millimetres, inches and as a percentage. Input the previously
measured filling level using the
[+] PLUS and
[-] MINUS
keys.
The filling level is specified in 1 millimetre steps. The inch,
percentage and milliampere units are calculated and displayed
for information.
The adjustment can only be made during operation. The tank
must be filled and the correct tank height must have been
entered so that the deviations remain small. Detailed instructions
can be found in Chapter 5 "Commissioning".
It can happen that a switching point is less than 4 mA or more than
20 mA by changing the span factor. If this happens, the relays do not
switch as desired and malfunctions can occur.
Check the current values in the menu items "Relays -> K1...K5 -> S-P On
and S-P Off". They must not be less than 4 mA and not more than
20 mA.
6.9.2
6.9.2
6.9.2
6.9.2
6.9.2 Settings menu Reset
Settings menu Reset
Settings menu Reset
Settings menu Reset
Settings menu Reset
With "Reset", you reset the correction factor for the sensor
adjustment back to the delivered condition. Select the symbol
[Y] - for "Yes, reset" and
[N] - for "No, do not reset".
Fig. 17:
"Change" display
Fig. 18:
"Reset" display