9
3.1
Completing an oil test using a Dräger Impactor or tube
The instrument displays the date and result from
the last oil Impactor or tube test.
To undertake a new test tap the result button.
Choose between using a Dräger oil Impactor or
Dräger chemical reagent tube in the test port.
For breathing-air tests the most commonly used
option is to select a Dräger oil Impactor.
Instructions for using the Dräger tube can be
found in appendix 1.
The instrument will now commence a 5-minute
purge sequence, during which time an airflow
will be passed through the test port. This clears
contaminants from previous tests and ensures
that a representative sample is being tested.
Once the purge has finished, the Dräger
Impactor/oil tube should be inserted into the
test port and tightened firmly; ensure the
Impactor has an intact protective seal. This seal
must remain in place for the duration of the
test. Press the arrow on the right hand side to
start the test.
At the end of the test the Impactor result screen
will be shown. The Impactor should now be
taken out of the test port, sticker removed and
evaluated.
To enter the results press the corresponding icon
on the screen.
Note:- Instructions on reading the Impactor can
be found in Appendix 2 at the back of the
manual.
The test result and date of test will then be
displayed on the home screen.