Compressed air  

2.

 

 Compressed air for BA normally originates from a compressor system. The maintenance, 
examination and testing of compressors should be carried out according to the manufacturer’s 
instructions. The siting of air inlets to compressors should follow the same principles as for 
fresh air hose. However, because compressors themselves can generate and concentrate a 
wide range of contaminants, you should take extra care in assuring air quality.  
 

3.

 

As the BA wearer’s life and health depend on the air supplied by the compressor, you should 
ensure that the air supplied meets the quality requirements in British Standard BS EN 12021 
Respiratory protective devices. Compressed air for breathing apparatus,* in addition to the 
pressure and airflow rate requirements of the BA manufacturer.  
 

4.

 

Compressors which are moved from site to site, such as those used by the emergency services 
or on construction sites, will require a higher standard of maintenance and should be sited so 
that the quality of air they provide is not compromised by nearby contaminants.  

 
* BS EN 12021 states: ‘Compressed air for breathing apparatus shall not contain any contaminants at 
a concentration which can cause toxic or harmful effects. In any event all contaminants shall be kept 
to as low a level as possible and shall be far below the national exposure limit. Combination effects of 
more than one contaminant shall be taken into account.’ (1999)Respiratory protective equipment at 
work Page 48 of 59  
 

Periodic testing of air quality  

5.

 

 The purpose of periodically testing air quality is to make sure that the control measures you 
have put in place are delivering the air quality required by BS EN 12021. You should base the 
frequency of such tests on a risk assessment, but they should take place at least every three 
months, and more often when the quality of air cannot be assured to these levels.  

 

6.

 

 As part of the risk assessment, if a mobile compressor is being used consideration should be 
given as to how often the air supply should be checked when the compressor is moved. Testing 
for these components may be carried out using any appropriate method, eg:  

simple colour change tubes; 
on-line gas testers; 
sample collection for laboratory analysis elsewhere.  

 

7.

 

 The supplier of your compressor or BA should be able to advise you on the best method for 
you. You should keep records of air quality tests for five years. 

 
 

© Crown copyright material is reproduced with the permission of the Controller of HMSO and Queen’s Printer 
for Scotland. 
 
Source Acknowledgement: 
HSG53 Respiratory Protective Equipment at Work ISBN 978 0717 6454 2 Health and Safety Executive 2013 
Contains public sector information licensed under the Open Government Licence v3.0

 

 
 

Содержание F6001

Страница 1: ...0 F6001 SAFE AIR TESTER OPERATING MANUAL FAC QAM 096 Issue D 03 21...

Страница 2: ...12 Appendix 2 Using the test port for other Draeger tubes 14 Appendix 3 Reading and disposal of Draeger Tubes and Oil Impactor 16 Appendix 4 Helpful Tips 21 Appendix 5 Air Quality Testing Why 23 Inter...

Страница 3: ...mpactors are supplied in packs of 10 and show the degree of oil contamination on a screen Calibration and Warranty Safe Air Testers leave our factory with a 12 month warranty and calibration certifica...

Страница 4: ...Draeger Impactor and tube test port Air Inlet F3002 HP Regulator optional extra Storage for manual and F3002 HP regulator hose Mains power adaptor TUB0053 8103560 Draeger Oil Impactors optional extra...

Страница 5: ...with this instrument Batteries must be installed prior to use unless it is the intention to power the unit solely from a mains supply To install the batteries lift the tester out of its case open the...

Страница 6: ...rial number and a tick or cross symbol will indicate whether it is within the recommended calibration period Press the arrow key to progress to the main menu The main menu has 3 options available from...

Страница 7: ...ration will operate for an extended period to stabilise the Oxygen cell You now have the option of entering a description for the test location if you wish to use numbers press the function button Pus...

Страница 8: ...hat the flow control valve is fully closed and then connect the F3002 high pressure regulator assembly to the cylinder or charging hose to be tested ensuring that the regulator has the correct type of...

Страница 9: ...ll be passed through the test port and across the electronic cells This clears contaminants from previous tests and ensures that a representative sample is being tested Once the purge has finished the...

Страница 10: ...fail against the standard displayed in the top right corner of the screen For water content on high pressure tests both the measured value in mg m3 and the atmospheric dewpoint are provided If the te...

Страница 11: ...n press the right hand arrow To delete a test highlight it then press the trash can button All tests can be deleted by selected the multiple trash can icon Note Once the instrument has saved 20 result...

Страница 12: ...Tester Results F4000 F6000 series software Then click on the File drop down menu and select import test results from F6000 via USB connection You will then be able to choose the location where you wou...

Страница 13: ...st due Test engineer and relevant address Note These details can be saved as default When complete you can save the test result file and if required print off a certificate In the UK Health and Safety...

Страница 14: ...e and time are set prior to the unit being supplied but they can be adjusted as necessary To select the units of measure and test standard press the right hand icon In this menu you can then choose da...

Страница 15: ...new test and after the unit had calibrated the oxygen cell and you have input the test location details you will have the option to select a tube test as shown press the button to proceed In this sub...

Страница 16: ...corner to proceed The test will proceed as described in section 3 and at the end of it you will be prompted to enter the result from the tube For the Nitrous fumes and Sulphur dioxide tubes enter the...

Страница 17: ...ickly identify minute quantities of oil For tests that pass with an oil concentration of less than 0 05mg m3 the screen will remain blank For tests above that read as follows The bottom line represent...

Страница 18: ...Using the Tube Tip Cutter place the ampoule section in the platform at the bottom Note Make sure the tube is against the back wall of the cutter Line up the black dot nearest the end of the tube with...

Страница 19: ...a slight discoloration Failed test Mineral oil The white crystals will show a light brown or darker discoloration Synthetic oil The white crystals will show a yellow discoloration Note best seen by c...

Страница 20: ...nvestigation will be required to determine the source of the contamination Disposal of Draeger Tubes When Draeger tubes have been used or unopened tubes have exceeded their expiry date they should be...

Страница 21: ...hand pump Break any ampoules where applicable Dispose of the tubes as stated in Methods 1 2 or 3 NOTE As an alternative a local authority approved waste disposal contractor can be employed to collect...

Страница 22: ...ic dewpoint The final test result for moisture will therefore display the limit for the sensor Make sure the system flow control valve is closed after each test Connecting an air supply to the Tester...

Страница 23: ...tor tube with the arrows facing outwards Never break the glass ampoule in the oil tube before the test Only use the detector tubes that the Tester is calibrated for Once used be aware that the tube ti...

Страница 24: ...t in air entering the filtration at too high a temperature this will cause the filtration to prematurely fail and pass excess levels of oil and water Malfunctioning dryers can disturb the oxygen conce...

Страница 25: ...onditions of usage and storage of any compressed air supply is not known the pressure dewpoint shall not exceed 11 C Dewpoint 50 F 67 PPM v v for SCBA use in extreme cold a dew point not to exceed 65...

Страница 26: ...ty is tested whenever it is first moved into a new position or prevailing wind conditions change The final decision on frequency of test is the responsibility of the Employer and needs to not only ref...

Страница 27: ...e condition of the equipment and details of any defect found including for canister or filter respirators the state of the canister and the condition of the filter for self contained compressed air ga...

Страница 28: ...quality of the air supplied Table A 2 provides a summary of the main elements associated with a compressor system for producing breathable air In addition to the careful and installation of the system...

Страница 29: ...quency of use When the air supply is from mobile compressors the employer should ensure that wherever a compressor is located the quality of air it supplies is not compromised by nearby contaminants I...

Страница 30: ...is not in use The adequacy of the accommodation will vary according to the quantity type and its use eg pegs labelled lockers shelves or containers etc The storage should be adequate to protect the PP...

Страница 31: ...compatible with the BA Never place the hose inlet near to potential sources of contamination eg vehicle exhausts Never use the equipment without the waist belt Never use a light duty airline hose whe...

Страница 32: ...to as low a level as possible and shall be far below the national exposure limit Combination effects of more than one contaminant shall be taken into account 1999 Respiratory protective equipment at...

Страница 33: ...32 Factair Ltd 49 Boss Hall Road Ipswich Suffolk IP1 5BN UK Tel Sales 44 0 1473 746400 Tel Hire 44 0 1473 746444 Fax 44 0 1473 747123 Email enquiries factair co uk www factair co uk...

Отзывы: