Factair F6001 Скачать руководство пользователя страница 17

16 

 

Oil Impactor 
 
The Oil Impactor’s protective seal must be kept in place for the duration of the test and only 
removed after the test has been completed. 
 

With  a  standard  measurement  range  of  0.1  to  1.0  mg/m

3

  the  Impactor  has  a  limit  of  detection  of 

0.05 mg/m

3

.    The  Impactor  can  detect  all  mineral  and  synthetic  oil  mists,  it  features  a  series  of  3 

horizontal lines, each formed by a series of precision manufactured nozzles.  These horizontal lines of 
nozzles are calibrated to a different concentration of oil.  When air is passed through these nozzles any 
oil present in the air is deposited on the glass plate prior to the air being exhausted by vents around its 
circumference, this allows the user to easily and quickly identify minute quantities of oil.   
 
For tests that pass with an oil concentration of less than 0.05mg/m

3

, the screen will remain blank.  For 

tests above that read as follows: 
 
The  bottom  line  represents  an  oil  concentration  of  0.1  mg/m

3

.    When  the  oil  deposited  forms  a 

continuous  line  then  the  concentration  is  in  excess  of  0.1  mg/m

3

.    The  middle  line  represents  a 

concentration of 0.5 mg/m

3

, again when the oil deposited forms a continuous line the concentration is 

in excess of 0.5 mg/m

3

.  The top line represents a concentration of 1.0 mg/m

3

, again when this forms a 

continuous line the concentration is in excess of 1.0 mg/m

3

.

 

     

 

 

 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 

 

 

* – The above pass and fail criteria is based on breathing-air standard EN12021 

Using the Draeger tube tip cutter 

 

Pass* - 
greater 
than 
0.05mg/m

3

 

but less 
than 
0.1mg/m

3

  

Pass* - 
greater 
than 
0.1mg/m

3

 

but less 
than 
0.5mg/m

3

  

Fail* - 
greater than 
0.5mg/m

3

 

but less than 
1.0mg/m

3

  

Fail* - 
greater 
than 
1.0mg/m

3

  

Содержание F6001

Страница 1: ...0 F6001 SAFE AIR TESTER OPERATING MANUAL FAC QAM 096 Issue D 03 21...

Страница 2: ...12 Appendix 2 Using the test port for other Draeger tubes 14 Appendix 3 Reading and disposal of Draeger Tubes and Oil Impactor 16 Appendix 4 Helpful Tips 21 Appendix 5 Air Quality Testing Why 23 Inter...

Страница 3: ...mpactors are supplied in packs of 10 and show the degree of oil contamination on a screen Calibration and Warranty Safe Air Testers leave our factory with a 12 month warranty and calibration certifica...

Страница 4: ...Draeger Impactor and tube test port Air Inlet F3002 HP Regulator optional extra Storage for manual and F3002 HP regulator hose Mains power adaptor TUB0053 8103560 Draeger Oil Impactors optional extra...

Страница 5: ...with this instrument Batteries must be installed prior to use unless it is the intention to power the unit solely from a mains supply To install the batteries lift the tester out of its case open the...

Страница 6: ...rial number and a tick or cross symbol will indicate whether it is within the recommended calibration period Press the arrow key to progress to the main menu The main menu has 3 options available from...

Страница 7: ...ration will operate for an extended period to stabilise the Oxygen cell You now have the option of entering a description for the test location if you wish to use numbers press the function button Pus...

Страница 8: ...hat the flow control valve is fully closed and then connect the F3002 high pressure regulator assembly to the cylinder or charging hose to be tested ensuring that the regulator has the correct type of...

Страница 9: ...ll be passed through the test port and across the electronic cells This clears contaminants from previous tests and ensures that a representative sample is being tested Once the purge has finished the...

Страница 10: ...fail against the standard displayed in the top right corner of the screen For water content on high pressure tests both the measured value in mg m3 and the atmospheric dewpoint are provided If the te...

Страница 11: ...n press the right hand arrow To delete a test highlight it then press the trash can button All tests can be deleted by selected the multiple trash can icon Note Once the instrument has saved 20 result...

Страница 12: ...Tester Results F4000 F6000 series software Then click on the File drop down menu and select import test results from F6000 via USB connection You will then be able to choose the location where you wou...

Страница 13: ...st due Test engineer and relevant address Note These details can be saved as default When complete you can save the test result file and if required print off a certificate In the UK Health and Safety...

Страница 14: ...e and time are set prior to the unit being supplied but they can be adjusted as necessary To select the units of measure and test standard press the right hand icon In this menu you can then choose da...

Страница 15: ...new test and after the unit had calibrated the oxygen cell and you have input the test location details you will have the option to select a tube test as shown press the button to proceed In this sub...

Страница 16: ...corner to proceed The test will proceed as described in section 3 and at the end of it you will be prompted to enter the result from the tube For the Nitrous fumes and Sulphur dioxide tubes enter the...

Страница 17: ...ickly identify minute quantities of oil For tests that pass with an oil concentration of less than 0 05mg m3 the screen will remain blank For tests above that read as follows The bottom line represent...

Страница 18: ...Using the Tube Tip Cutter place the ampoule section in the platform at the bottom Note Make sure the tube is against the back wall of the cutter Line up the black dot nearest the end of the tube with...

Страница 19: ...a slight discoloration Failed test Mineral oil The white crystals will show a light brown or darker discoloration Synthetic oil The white crystals will show a yellow discoloration Note best seen by c...

Страница 20: ...nvestigation will be required to determine the source of the contamination Disposal of Draeger Tubes When Draeger tubes have been used or unopened tubes have exceeded their expiry date they should be...

Страница 21: ...hand pump Break any ampoules where applicable Dispose of the tubes as stated in Methods 1 2 or 3 NOTE As an alternative a local authority approved waste disposal contractor can be employed to collect...

Страница 22: ...ic dewpoint The final test result for moisture will therefore display the limit for the sensor Make sure the system flow control valve is closed after each test Connecting an air supply to the Tester...

Страница 23: ...tor tube with the arrows facing outwards Never break the glass ampoule in the oil tube before the test Only use the detector tubes that the Tester is calibrated for Once used be aware that the tube ti...

Страница 24: ...t in air entering the filtration at too high a temperature this will cause the filtration to prematurely fail and pass excess levels of oil and water Malfunctioning dryers can disturb the oxygen conce...

Страница 25: ...onditions of usage and storage of any compressed air supply is not known the pressure dewpoint shall not exceed 11 C Dewpoint 50 F 67 PPM v v for SCBA use in extreme cold a dew point not to exceed 65...

Страница 26: ...ty is tested whenever it is first moved into a new position or prevailing wind conditions change The final decision on frequency of test is the responsibility of the Employer and needs to not only ref...

Страница 27: ...e condition of the equipment and details of any defect found including for canister or filter respirators the state of the canister and the condition of the filter for self contained compressed air ga...

Страница 28: ...quality of the air supplied Table A 2 provides a summary of the main elements associated with a compressor system for producing breathable air In addition to the careful and installation of the system...

Страница 29: ...quency of use When the air supply is from mobile compressors the employer should ensure that wherever a compressor is located the quality of air it supplies is not compromised by nearby contaminants I...

Страница 30: ...is not in use The adequacy of the accommodation will vary according to the quantity type and its use eg pegs labelled lockers shelves or containers etc The storage should be adequate to protect the PP...

Страница 31: ...compatible with the BA Never place the hose inlet near to potential sources of contamination eg vehicle exhausts Never use the equipment without the waist belt Never use a light duty airline hose whe...

Страница 32: ...to as low a level as possible and shall be far below the national exposure limit Combination effects of more than one contaminant shall be taken into account 1999 Respiratory protective equipment at...

Страница 33: ...32 Factair Ltd 49 Boss Hall Road Ipswich Suffolk IP1 5BN UK Tel Sales 44 0 1473 746400 Tel Hire 44 0 1473 746444 Fax 44 0 1473 747123 Email enquiries factair co uk www factair co uk...

Отзывы: