Factair F3000 Скачать руководство пользователя страница 2

 

 

CONTENTS 

 

   

 

 

 

 

 

Page(s) 

 

 

INTRODUCTION 

 

 

 

 

 

 

 

1

 

 
 
 

 

BATTERY AND POWER SUPPLY 

 

 

 

 

2

 

   

 

 
 

3   

 

OPERATING INSTRUCTIONS 

 

 

 

 

 

 

 

 

3.1 

Testing of low-pressure breathing airlines and mobiles  

3-7 

  3.2 

Testing high-pressure cylinders   

 

 

 

7-11 

 
 
 

4

 

 

APPENDICES

 

 

Appendix 1 - Helpful tips   

 

 

 

 

 

12 

Appendix 2 - Reading Draeger Detector Tubes and the safe 

 

disposal thereof 

 

 

 

 

 

13 

Appendix 3 -  Disposal of Draeger-Tubes  

 

 

 

14 

Appendix 4 -  Establish Pressure Dewpoint 

 

 

 

15

 

Appendix 5 - Pressure Dewpoint Graph   

 

 

 

16 

 

 

 

 

Содержание F3000

Страница 1: ...F3000 SAFE AIR TESTER OPERATING INSTRUCTIONS Factair Ltd 49 Boss Hall Road Ipswich Suffolk IP1 5BN Tel no 44 0 1473 746400 Fax no 44 0 1473 747123 Email enquiries factair co uk Website www factair co uk ...

Страница 2: ...re breathing airlines and mobiles 3 7 3 2 Testing high pressure cylinders 7 11 4 APPENDICES Appendix 1 Helpful tips 12 Appendix 2 Reading Draeger Detector Tubes and the safe disposal thereof 13 Appendix 3 Disposal of Draeger Tubes 14 Appendix 4 Establish Pressure Dewpoint 15 Appendix 5 Pressure Dewpoint Graph 16 ...

Страница 3: ...ng the unit Calibration and Warranty Safe Air Testers leave our factory with a 12 month warranty and calibration certificate Our standard turnaround on annual calibration is 10 working days providing there is no major damage that requires extensive rebuild Inside the carry case there is a laminated sheet listing the detector tubes that your Safe Air Tester is calibrated for and information on how ...

Страница 4: ...peat tests will be permitted The screen will display Battery low Replace batteries To replace the batteries i switch off the tester ii turn the unit onto its back iii release the battery tray lid situated in the base of the tester iv remove old batteries and refit new batteries ensuring that polarity is correct ie to and to v dispose of used batteries in a responsible manner Note The unit will aut...

Страница 5: ...ssure 10 bar Press enter IMPORTANT If the unit is overpressurised it will automatically shut down and display an overpressure warning it will then need to be returned to the manufacturer for inspection and resetting Note The above service is chargeable iv Adjust the system flow control valve so the normal flow required for the face piece shows on the flowmeter best achieved by raising beyond the s...

Страница 6: ...nge by means of the up down scroll buttons HI 50 2000 mg m for standard airline use LO 5 200 mg m for airlines incorporating a desiccant dryer Press enter vi Select oil time Screen will display Preset times Add custom time These are selectable by using the up down scroll buttons Press enter Select according to the following requirements vii Preset times The following common times are available wit...

Страница 7: ...o the test times chart provided using the column marked for the F3000 If the required test time is one of the quoted preset times then select using the up down scroll buttons Press enter Note for a test time not listed in the preset menu select Add custom time at vi Press enter viii Add custom time Using left right scroll buttons move the cursor over the time to be altered and set the appropriate ...

Страница 8: ...ectly with the arrows pointing outwards Press enter xii Testing time will be displayed as a countdown together with airline temperature ambient temperature oxygen content and airline pressure The status of the individual tests is indicated on the left hand side of the display Black test active clear test completed The unit can be left until testing time has elapsed xiii Display will read Testing c...

Страница 9: ...ers i Turn on power by pressing and holding the ON OFF button on the keypad for approximately 2 5 seconds The unit will display the model and serial number Press enter to confirm the date of the last calibration and the battery condition Note It is recommended that your F3000 is calibrated within a 12 month period ii If the battery condition is OK Press enter iii Ensure the system flow control val...

Страница 10: ...t above the flowmeter then close the system flow control valve Use the minimum amount of air to preserve cylinder capacity Note If there is a serious odour problem fail the sample at this point of the test Running the tester with seriously contaminated air can damage the unit Press enter Detector Tube Set up Note CO and CO2 are set Water and oil channels are variable and must be set as below v Sel...

Страница 11: ...n accept the default time of 15 minutes by pressing enter If you can determine with a high degree of certainty the specific compressor lubricant being used then refer to the test times chart provided using the column marked for the F3000 If the required test time is one of the quoted preset times then select using the up down scroll buttons Press enter Note for a test time not listed in the preset...

Страница 12: ... tubes with the special cutter provided and ensuring that no glass shards are present in the glands insert the tubes into the relevant ports and tighten the gland covers Always check to ensure tubes are inserted correctly with the arrows pointing outwards Press enter xii Testing time will be displayed as a countdown together with airline temperature ambient temperature oxygen content and airline p...

Страница 13: ... hold the ON OFF button for approximately 2 5 seconds Note The tester will automatically switch off if not used after a period of 1 hour 11 APPENDIX 1 The following information is only for EU member states The use of the symbol indicates that this product may not be treated as household waste By ensuring this product is disposed of correctly you will help prevent potential negative consequences fo...

Страница 14: ...d to the tube whilst rotating for scoring General Make sure the system flow control valve is closed after each test Note Connecting an air supply to the tester with the system flow regulator in the open position can damage the unit Never connect to a non regulated supply from an HP cylinder or compressor Maximum inlet pressure to the tester is 10 bar If exposed to an overpressure the tester will d...

Страница 15: ...OINT TO BE 5 C BELOW THE LIKELY LOWEST TEMPERATURE WHERE THE LIKELY LOWEST TEMPERATURE IS NOT KNOWN THE PRESSURE DEWPOINT SHOULD NOT EXCEED 11 C FOR HIGH PRESSURE CYLINDERS THE FOLLOWING UPPER LIMITS APPLY 40 TO 200 BAR 50MG M ABOVE 200BAR 35MG M FOR HIGH PRESSURE CYLINDER CHARGING COMPRESSORS THE UPPER LIMIT 25MG M A reddish brown discoloration will show the extent of the water content which is r...

Страница 16: ...ration OR 2 Place the tubes in a sharps or glass bin wearing protective gloves and safety spectacles Dispose of the bin via the company s normal hazardous waste disposal method s ie landfill or incineration OR 3 Place the tubes in a sharps or glass bin wearing protective gloves and safety spectacles Dispose of the container via incineration Unused Tubes Open the Draeger tube at both ends using the...

Страница 17: ... the actual reading ie a tube reading of 150 mg m becomes 1500 mg m To establish the pressure dewpoint reading to be entered on the results pad us the Pressure Dewpoint Graph for airline systems printed on the reverse of the results pad and apply the following procedures Take the mg m reading from the bottom axis of the graph and extend until it crosses the line for the airline pressure of the sys...

Страница 18: ... m3 of Water 80 70 60 50 40 30 20 10 0 0 5 10 30 50 70 100 150 200 80 70 60 50 40 30 20 10 0 1 Bar 3 Bar 5 Bar 7 Bar 10 Bar To establish pressure dewpoint take the point where tube reading in mg m3 intersects system pressure and read off pressure dewpoint from vertical scale FAC QAM 075 06 05 ISSUE A ...

Страница 19: ......

Страница 20: ...s Hall Road Ipswich Suffolk IP1 5BN Tel no 44 0 1473 746400 Fax no 44 0 1473 747123 Email enquiries factair co uk Website www factair co uk System Flowmeter Carbon Monoxide Carbon Dioxide Oil Water Electronic multifunction display System Flow Regulator Keypad Power On Off Air Inlet ...

Отзывы: