background image

18

The

SpO

2

and

Pulse

Rate
are not

displaye

d stably

1. The finger
is not placed
inside deep
enough.
2. The finger
is shaking or
the patient is
moving.

1. Place the
finger properly
and try again.
2. Let the
patient keep
calm

The

device
can not

be

turned
on

1. 1. The
battery is
drained away
or almost
drained away.
2. The battery
is installed
incorrectly.
3.The
malfunction of
the device.

1.Please
change
batteries.
2.Please Install
the battery
again.
3. Please
contact the
local service
center.

The

display
is off

suddenl

y

1. The
device

s

malfunction.
2. The battery

is drained
away or almost
drained away.

1. Please
contact the
local service
center.
2.Please
change
batteries.

19

9 Key of Symbols

Signal

Description

Warning

See User Manual

%SpO

2

The pulse oxygen saturation(%)

PRbpm

Pulse rate (bpm)

Low-voltage

menu button/power
button/function button

Type BF

SN

Serial number

1. the finger clip falls off ( no
finger inserted)]
2. Probe error
3. Signal inadequacy indicator

Содержание CMS-50D Plus

Страница 1: ......

Страница 2: ...the Pulse Oximeter s features and requirements main structure functions specifications correct methods for transportation installation usage operation repair maintenance and storage etc as well as th...

Страница 3: ...evice ceaselessly especially for the microcirculation barrier patients It is recommended that the sensor should not be applied to the same finger for over 2 hours For the special patients there should...

Страница 4: ...s Check the main unit and all accessories periodically to make sure that there is no visible damage that may affect patient s safety and monitoring performance about cables and transducers It is recom...

Страница 5: ...r is not permitted Refer to User Manual in the relative chapter 7 1 for 3 instructions of cleaning and disinfection Do not have the oximeter immerged in liquid When it needs cleaning please wipe its s...

Страница 6: ...able It is only necessary for patients to put one of his fingers into a probe for diagnosis and a display screen will directly show the measured value of pulse oxygen saturation with the high veracity...

Страница 7: ...sitive element information acquired through which will be shown on screen through treatment in electronic circuits and microprocessor Figure 1 7 4 Technical specifications 4 1 Main performance A SpO2...

Страница 8: ...nm 6 75mW 5 Installation 5 1 View of the front panel Figure 2 front view 5 2 Battery installation A Refer to Figure 3 and insert the two AAA size batteries properly in the right direction 9 B Replace...

Страница 9: ...tion The icon of SpO2 is a program for receiving real time data which is shown as figure 5 the icon of SpO2 Review is a program for receiving stored data which is shown as figure 6 11 a Please connect...

Страница 10: ...nds and then the screen will be automatically shut down only a flashing yellow dot appear on screen If short press the button at this moment the sign Recording would appear on the screen and then the...

Страница 11: ...sensor should not be used at a location 15 or limb tied with arterial canal or blood pressure cuff or receiving intravenous injection E Do not fix the SpO2 sensor with adhesive or else it may result i...

Страница 12: ...n shows C Take out the battery if leave the equipment unused for long time D The device needs to be calibrated once a year or according to the calibrating program of hospital It also can be performed...

Страница 13: ...vice 1 Please change batteries 2 Please Install the battery again 3 Please contact the local service center The display is off suddenl y 1 The device s malfunction 2 The battery is drained away or alm...

Страница 14: ...ion Specification Information Display Mode The Pulse Oxygen Saturation SpO2 2 digit digital OLED display Pulse Rate PR 3 digit digital OLED display Pulse Intensity bar graph bar graph OLED display SpO...

Страница 15: ...Moving calculate the Average pulse rate every 4 cardio beat s cycle The deviation between average value and true value does not exceed 1 Safety Type Interior Battery Type 23 Pulse Intensity Range Con...

Страница 16: ...24 Dimensions 58 5 L 31 W 32 H mm Weight About 50g with the batteries...

Страница 17: ...File Ver 0 9 Release Date March 2009...

Отзывы: