Communication with Various Vision Test Devices
ERK-7800 could be connected a various optometry devices with a single network. Lensmeter has
not yet been connected because it’d be applied in the near future. Here is the list to be connected.
Chart Projector: ECP-5400, LCD CHART
Auto Ref / Keratometer: ERK-5400, ERK-7800
7.1
7.1
7.1
7.1
Option setting for Chart Projector (ECP-5400)
To use EDR 7800 proper, Chart Projector (ECP-5400) must be installed. In case Chart Projector is
installed, just follow as below to choose the options
SYSTEM CONFIG – page 7 - [EXTERNAL INSTRUMENTS]
CP CONNECTION: Must choose SERIAL. If choose NONE, ECP-5400 does not run.
To set up the above option, enter the [SYSTEM CONFIG] mode through [MENU] mode, and move
on to page 7 by pressing [#7] button. Then press [OK] button to save it after setting all
[Figure 7-1] Chart projector setup on SYSTEM CONFIG – page 7
7.2
7.2
7.2
7.2
Option setting for Auto Ref/Keratometer (ERK-7800)
It’s recommended to install ERK-7800 to make full use of the function of EDR 7800
Although there is ‘RK: COPY KERATO’ menu at [EXTERNAL INSTRUMENTS] on SYSTEM
CONFIG - page 7 -, it’s useless now because ERK-7800 does not transmit KERATO data. It is only
for later use when it is available.
7.3
7.3
7.3
7.3
Load and Save the data of Auto Ref / Keratometer (ERK-7800)
Transmit the test results of the Auto Ref / Keratometer (ERK-7800) to EDR 7800
①
Measure the patient’s eye by auto ref/keratometer
②
The test results are transmitted to EDR-7800 from ERK-7800, by pressing PRINT button
on ERK-7800.