9. Setting up the chair for use

The product is delivered completely assembled in a 
carton. Before using the first time, the chair first must 
be checked over for transport damage. Settings and 
adjustments may be done only by a qualified specialist. 
The hydraulic system may need to be vented before 
use because an air pocket may have formed during the 
transport. See Section 13.2 on page 12.

9.1 Position of the footrests

Rotate the footrests (1) all the way to the front until 
they latch in position. Next unfold the foot plates (2).
The foot plates may need to be height adjusted.

9.2 Adjusting the foot plate height

The foot plates are height-adjustable to four
positions. Remove the screws from the side at the
footrests using an SW 5 Allen key. Select the
appropriate level (1-4) and once again tighten the
screws at the foot plates.
Be sure that the screws are firm!

  

9.3 Installing, adjusting the headrest 

Installation:

Flip the levers (A) upwards 

and insert the tubes of 

the headrest unit into the 

openings.

Push the tubes down until 

the push-buttons (C) click 

in the holes (D).

Next push the button (C) 

all the way in and push the 

tube to the end and press 

the levers (A) all the way 

down.

Setting the height and depth:

To set the headrest height, flip both levers (A) 
upwards and push the head rest to the appropriate 
height.

The head rest can be

steplessly adjusted (ca.

4.72") until the button (D) 

clicks into the hole  

(end-position).

After setting, press both

levers (A) down again. 

By removing the clips 

(C), the headrests can be 

removed for cleaning or for 

replacing.

9.4 Setting the inclination

The backrest may be steplessly inclined back up to
approx. 37°. Here the seat inclines up to 12°. Next
pull the locking levers (A) vertically upward and
push the backrest (B) downward until the desired
position is reached. Next press the locking levers
(C) all the way down.

When adjusting the backrest 

inclination, be sure that no body 

parts or objects get pinched and that 

the armrests are latched in position!

Содержание EZ352

Страница 1: ...30 18 EZ Way Inc PO Box 89 Clarinda IA 51632 1 800 627 8940 www ezlifts com Operator s Instructions Dignity Hydraulic Height Adjustable Reclining Shower Chair WARNING Cancer and Reproductive Harm www...

Страница 2: ...6 Safety advisories for use 4 7 Product specifications 5 7 1 Delivery scope 5 7 2 Technical specifications 6 8 Spare parts accessories 6 9 Setting up the chair for use 7 9 1 Position of the footrests...

Страница 3: ...h a relative humidity of 30 50 The product may be used outside of these ranges but only for a short period of time Avoid heat sources impacting the product for longer time periods see safety advisory...

Страница 4: ...al company or environmental authorities for further information 6 Safety advisories for use Abide by the following safety advisories when using the chair Danger of slipping Before use check to be sure...

Страница 5: ...clean it with hot water Allow the surface of the chair to first cool off before use When the chair is to be used by patients with insensitive skin persons who cannot feel temperature rises and or hav...

Страница 6: ...dal activation force for 330 lbs load ca 66 14 lbs Skin compatibility As to DIN EN 10993 5 Inflammability As to DIN EN 1021 2 Dimensions for model w head rest roller Design 1 Flame retardant article m...

Страница 7: ...ng the headrest Installation Flip the levers A upwards and insert the tubes of the headrest unit into the openings Push the tubes down until the push buttons C click in the holes D Next push the butto...

Страница 8: ...gle can be adjusted Warning Danger of pinching B between the headrest and backrest 9 6 Adjusting the seat height The chair s seat height is done in steps via the hydraulic unit by activating the foot...

Страница 9: ...onto the chair seat must be avoided The patient must sit as much as possible over the entire seat surface and not only on the seat edge When transporting a person the surface of the seat and the backr...

Страница 10: ...ever upwards with your foot 2 10 6 Removing the seat upholstery The seat with I and without O hygiene opening the leg pad K and the backrest upholstery H can be removed from their mountings in the cha...

Страница 11: ...C must be put back on the shower seat 11 Troubleshooting Chair malfunctions faults Fault Possible cause Remedy Armrests cannot be swiveled out Caking up in the latching latching pins Clean check regu...

Страница 12: ...sually the unit can remain in the frame Clean the piston rod 12 Care 12 1 Caring for the chair In general the wheelchair can be cleaned using a mild conventional cleaning and disinfecting agent Next s...

Страница 13: ...ew on the unit for the purpose of checking the oil or for filling Venting To vent the hydraulic unit the unit must be removed To do this remove the keyrings at points A and B and then pull out both bo...

Страница 14: ...tion of the chair When making repairs use only original spare parts from REBOTEC and LINAK Observe all advisories concerning guarantee and warranty Once a complete maintenance or repair of the chair i...

Страница 15: ...15...

Страница 16: ...oned by the use loss of use or misuse of the products or by the failure of the products to conform to the warranty set forth above This warranty becomes void if the product shows evidence of mishandli...

Отзывы: