background image

13

  

13.1 Maintenance on the chair

Regular safety-technical checks are necessary:

• 

Check the product for visible damage (cracks, 
fractures, loose parts, instability). 

• 

All screwed connections for tightness 
(especially the castor wheels and frame). 

• 

Function of the armrests and latching pins. 

• 

Function of the footrests (replace defective foot 
plates). 

• 

Rolling ability, check function of the wheels 
(replace defective wheels). 

• 

Function of the castor locking brakes (replace 
defective brakes). 

• 

Function of the 

headrest

• 

Function of the chair inclination. 

• 

Check accessories for inadequacies.

• 

In the event of any inadequacies, do not put the 
product to use and do not carry out any repairs 
on your own.  
Contact EZ Way at 

(800) 627-8940.

13.2 Maintenance on the hydraulic unit

→ 

Hydraulic cylinder

Regular safety-technical checks are necessary:

• 

There is a risk of slipping and falling if hydraulic 
luid leaks out!

• 

Visually check the unit for leaks every time 
before use.

• 

In the event of any inadequacies, do not put the 
product to use and do not carry out any repairs 
on your own.

• 

Contact EZ Way at 

(800) 627-8940. 

→ 

Hydraulic unit

The hydraulic unit is a closed system and thus no
maintenance is needed, Note the following:

• 

The unit does not have an oil fill plug.

• 

The oil fill level cannot be checked.

• 

Do not open any screw on the unit for the 
purpose of checking the oil or for filling. 
 

• 

Venting:

To vent the hydraulic unit, the unit must be
removed. To do this, remove the keyrings at points
A and B, and then pull out both bolts from the
holder, one after the other.

To completely vent the unit, clamp it in the vertical
position with the piston extended out and wait ten
minutes. Suitable to use as a clamping point is the
port of the hydraulic unit at Point A.
Be sure that the strike plate does not get clamped
because it may be bent. First push it to the side.

Proceed with venting as follows:
1. Move the piston rod completely out by pressing
the pedal multiple times, and activate the pedal
multiple times after the end of the stroke is
reached. This will cause the unit to vent out by
itself via an integrated venting valve.

2. Drive entire hydraulic system back.

3. Plan maintenance time.

Repeat the process several times, or until the unit
reacts immediately to the pedal movement and the
system is rigid. Then the unit is vented.

Warning: There is a danger of 

injury from slipping and pinching!

• 

Replacing the seals:

Due to aging of the seal material, the seals must
be replaced about every 6 years.
Contact EZ Way.

13.3 Fire protection – hydraulic system

Combustible hydraulic fluid is used. If leaks occur
from the unit and contact is made between the fluid
and an open flame or a hot surface, then the fluid
could ignite.
It must first be assured that fires can be put out
immediately with proper extinguishers.

Содержание EZ352

Страница 1: ...30 18 EZ Way Inc PO Box 89 Clarinda IA 51632 1 800 627 8940 www ezlifts com Operator s Instructions Dignity Hydraulic Height Adjustable Reclining Shower Chair WARNING Cancer and Reproductive Harm www...

Страница 2: ...6 Safety advisories for use 4 7 Product specifications 5 7 1 Delivery scope 5 7 2 Technical specifications 6 8 Spare parts accessories 6 9 Setting up the chair for use 7 9 1 Position of the footrests...

Страница 3: ...h a relative humidity of 30 50 The product may be used outside of these ranges but only for a short period of time Avoid heat sources impacting the product for longer time periods see safety advisory...

Страница 4: ...al company or environmental authorities for further information 6 Safety advisories for use Abide by the following safety advisories when using the chair Danger of slipping Before use check to be sure...

Страница 5: ...clean it with hot water Allow the surface of the chair to first cool off before use When the chair is to be used by patients with insensitive skin persons who cannot feel temperature rises and or hav...

Страница 6: ...dal activation force for 330 lbs load ca 66 14 lbs Skin compatibility As to DIN EN 10993 5 Inflammability As to DIN EN 1021 2 Dimensions for model w head rest roller Design 1 Flame retardant article m...

Страница 7: ...ng the headrest Installation Flip the levers A upwards and insert the tubes of the headrest unit into the openings Push the tubes down until the push buttons C click in the holes D Next push the butto...

Страница 8: ...gle can be adjusted Warning Danger of pinching B between the headrest and backrest 9 6 Adjusting the seat height The chair s seat height is done in steps via the hydraulic unit by activating the foot...

Страница 9: ...onto the chair seat must be avoided The patient must sit as much as possible over the entire seat surface and not only on the seat edge When transporting a person the surface of the seat and the backr...

Страница 10: ...ever upwards with your foot 2 10 6 Removing the seat upholstery The seat with I and without O hygiene opening the leg pad K and the backrest upholstery H can be removed from their mountings in the cha...

Страница 11: ...C must be put back on the shower seat 11 Troubleshooting Chair malfunctions faults Fault Possible cause Remedy Armrests cannot be swiveled out Caking up in the latching latching pins Clean check regu...

Страница 12: ...sually the unit can remain in the frame Clean the piston rod 12 Care 12 1 Caring for the chair In general the wheelchair can be cleaned using a mild conventional cleaning and disinfecting agent Next s...