3

1. General advisories

1.1 Introduction

In this booklet you will find information and 
advisories for start-up, use, operation, maintenance 
and care of the Dignity Hydraulic Height 
Adjustable Reclining Shower Chair.
Carefully read the Operator’s Instructions before 
using the product. Abide by the safety instructions 
and act accordingly.

1.2 Symbols used

Read the

instructions!

Year of

Manufacture

Product complies to the Medical

Products Directive 93/42/EEC

Caution!

Read the safety advisories

Attention!

Advisory

Important

Information!

Care

advisory

Maintenance

advisory

Danger of

hand injury

Stepping on

surface

forbidden

330 lbs.

Caution!

Abide to

max. user

weight or

load limit

330 lbs.

1.3 Intended proper use

• 

The shower chair is for homecare and 
institutional purposes. It is to be used 
exclusively for assistance when going to the 
toilet, showering or when transferring a person 
indoors.

• 

Rolling over a toilet is not possible!

• 

The shower room/stall should be constructed so 
that it is possible to travel into it with the chair. 
The chair is not suitable for extended periods of 
sitting and for long patient transfer distances.

• 

Persons with limited perception and senses, 
joint damage in the arms, and inabilities to sit 
may not use this wheelchair. 

• 

Maximum user weight is ca. 331 lbs.

1.4 Use and storage

• 

The product’s temperature range for use must be 
between 

+41°F and +104°F

. Store the product 

in a dry room with a relative humidity of  

30-50%

. The product may be used outside of 

these ranges, but only for a short period of time.

• 

Avoid heat sources impacting the product 
for longer time periods (see safety advisory: 
“Caution: danger of injury”).

• 

The chairs may not be used in swimming 
pools or sauna areas because the surface may 
change colors, and limitations with respect to 
mechanical stability may result.

1.5 Equipment

• 

The chairs are made of a plastic frame mounted 
on four 5-inch swivel castors equipped with 
locking brakes. The armrests can be swiveled 
and have a latching mechanism.

• 

The footrests are height-adjustable and may 
be removed or folded away. The headrests are 
height-adjustable and can be rotated.

• 

The chair’s inclination adjustment for the 
patient’s back position is controlled by a lever 
unit. 

• 

With the hydraulic lifting unit, the patient can 
be lifted approx. 15.7".

• 

Depending on the equipment, a PUR pad is 
included for transport purposes or a PUR pad 
with hygiene opening (including commode 
pan).

• 

Because of the product’s design and purpose, 
disassembly for transport and storing has not 
been planned and is not necessary. 

2. Warranty

We grant a 

warranty of 5 years

 on the product. 

The warranty is valid as long as the product is used 
in accordance to the operational requirements and 
within the general terms and conditions stated in 
this manual. 
Unauthorized modifications done to the  product 
without the manufacturer’s permission automatically 
leads to nullification of the warranty on the product.

The delivery date written on the 

REBOTEC- delivery note applies as the 

valid start date of the warranty period.

Содержание EZ352

Страница 1: ...30 18 EZ Way Inc PO Box 89 Clarinda IA 51632 1 800 627 8940 www ezlifts com Operator s Instructions Dignity Hydraulic Height Adjustable Reclining Shower Chair WARNING Cancer and Reproductive Harm www...

Страница 2: ...6 Safety advisories for use 4 7 Product specifications 5 7 1 Delivery scope 5 7 2 Technical specifications 6 8 Spare parts accessories 6 9 Setting up the chair for use 7 9 1 Position of the footrests...

Страница 3: ...h a relative humidity of 30 50 The product may be used outside of these ranges but only for a short period of time Avoid heat sources impacting the product for longer time periods see safety advisory...

Страница 4: ...al company or environmental authorities for further information 6 Safety advisories for use Abide by the following safety advisories when using the chair Danger of slipping Before use check to be sure...

Страница 5: ...clean it with hot water Allow the surface of the chair to first cool off before use When the chair is to be used by patients with insensitive skin persons who cannot feel temperature rises and or hav...

Страница 6: ...dal activation force for 330 lbs load ca 66 14 lbs Skin compatibility As to DIN EN 10993 5 Inflammability As to DIN EN 1021 2 Dimensions for model w head rest roller Design 1 Flame retardant article m...

Страница 7: ...ng the headrest Installation Flip the levers A upwards and insert the tubes of the headrest unit into the openings Push the tubes down until the push buttons C click in the holes D Next push the butto...

Страница 8: ...gle can be adjusted Warning Danger of pinching B between the headrest and backrest 9 6 Adjusting the seat height The chair s seat height is done in steps via the hydraulic unit by activating the foot...

Страница 9: ...onto the chair seat must be avoided The patient must sit as much as possible over the entire seat surface and not only on the seat edge When transporting a person the surface of the seat and the backr...

Страница 10: ...ever upwards with your foot 2 10 6 Removing the seat upholstery The seat with I and without O hygiene opening the leg pad K and the backrest upholstery H can be removed from their mountings in the cha...

Страница 11: ...C must be put back on the shower seat 11 Troubleshooting Chair malfunctions faults Fault Possible cause Remedy Armrests cannot be swiveled out Caking up in the latching latching pins Clean check regu...

Страница 12: ...sually the unit can remain in the frame Clean the piston rod 12 Care 12 1 Caring for the chair In general the wheelchair can be cleaned using a mild conventional cleaning and disinfecting agent Next s...

Страница 13: ...ew on the unit for the purpose of checking the oil or for filling Venting To vent the hydraulic unit the unit must be removed To do this remove the keyrings at points A and B and then pull out both bo...

Страница 14: ...tion of the chair When making repairs use only original spare parts from REBOTEC and LINAK Observe all advisories concerning guarantee and warranty Once a complete maintenance or repair of the chair i...

Страница 15: ...15...

Страница 16: ...oned by the use loss of use or misuse of the products or by the failure of the products to conform to the warranty set forth above This warranty becomes void if the product shows evidence of mishandli...

Отзывы: