34
Eyecon 9420 & 9430 Instruction Manual
Settings menu
Validation
If you select the
Validation
mode, you can use the choices on the right hand screen to
set the total length of the barcode and then the offsets and lengths of each part of the
barcode:
Rx#
,
Drug Number
and
Quantity
. This is illustrated by the example below
Refer to the screen below as you move through determining and setting the values.
To determine the values for each section, you will need one of the barcodes that will be
used.
1.
Tap
Validation Test Mode
in the right pane.
The Vial Barcode Test Mode Screen appears.
Validation Barcode Length = Overall barcode length
Example = 22
Offset: 0
Offset: 7
Offset: 18
Rx#
Length: 7
Drug Number
Length: 11
Qty.
Length: 4
1
2
3
4
5
6
7
8
9
10
11 12
13 14 15 16
17 18 19
20 21 22
7 7 7 7 7 7 7 1 2 3 4 5 6 7 8 9 0 1 0 0 5 0