TKG250 Ultrasonic Thickness Gauge
69
Gain-Manual or Automatic, Gain can be adjusted automatically where the gauge will set
the optimum range or manually, where the operator changes the range in 1dB steps. To
select Auto, use the left or right arrow keys and highlight gain on the bottom right side of
the display, use the up and down arrows to select Manual for desired setting or F2 for
AGC. The Wave is defaulted to Auto gain where Automatic Gain Control (AGC) is on.
Rectify-RF, Full Wave, Half+ and Half – Use the left or right arrow keys to highlight
RECT. for rectify the up or down arrows to select between RF, Full Wave, Half Wave +
or Half Wave -. The default setting is Full Wave Rectified.
Blanking
To adjust for main bang blank or blank after first received echo. Use the arrow keys to
highlight “Blanking”, select main bang blank or blank after first received echo (in Echo-
to-Echo mode only), press Menu/OK. Use the left or right arrows to adjust the blanking
accordingly to eliminate noise or mode converted echoes. The Default setting is Main
Bang Blank on. Blank after first received echo is only enabled once Echo-to-Echo is
turned on.