Extech Instruments
40 TKG250- EN V2.4 11/13
reading will be stored at the ID location [Linear], [Row], [Col] without notes.
(See: Directory selection for file type details.)
Using the Freeze Option
You can freeze your data via the
Freeze
option.
To use the
Freeze
option, follow these steps:
1.
From the
Measurement
mode screen, press
[F2]
(Freeze)
.
The following screen comes up after pressing
[F2]
(Freeze)
whereby showing the
flag “
Freeze
” under the thickness reading. You are now in
Freeze
mode. The
[F2]
selection also changes from
Freeze
to
Unfreeze
.
2.
Press
[F1] (Save)
to save the reading. (See Using the Save Option for more
information.)
3.
Press
[F2] (Unfreeze)
to disable the
Freeze
mode
4.
Press
[F3] (Dir)
to perform file operations. (See Using the Directory Option for
more information.)
5.
While in
Freeze
mode the menu options are disabled. The display will prompt
you to unfreeze the gauge.