Datalogger 

Types of Data Recording 

 

Manual Datalogging:

 Manually log up to 99 readings onto an SD card via push-key press. 

 

Automatic Datalogging:

 Automatically log data onto an SD memory card where the number of 

data points is limited only by the card size. Readings are logged at a rate specified by the user. 

SD Card Information 

 

Insert an SD card (from 1G size up to 16G) into the SD card slot at the bottom of the meter. 
With regard to orientation, the card should be inserted with the front of the card (label side) 
facing toward the rear of the meter. 

 

If the SD card is being used for the first time it is recommended that the card be formatted and 
the logger’s clock set to allow for accurate date/time stamping during datalogging sessions. 
Refer to the Setup Mode section for SD card formatting and time/date setting instructions. 

 

European and USA numerical formats differ. The data on the SD card can be formatted for 
either format. The meter defaults to USA mode where a decimal point is used to separate units 
from tenths, i.e. 

20.00

. The European format uses a comma, i.e. 

20,00

. To change this setting, 

refer to the Setup Mode section. 

Manual Datalogging 

In the manual mode the user presses the LOG key to manually log a reading onto the SD card. 

1. 

Set the sampling rate to ‘0’ seconds. 

2. 

Press and hold the LOG key for at least 1.5 seconds; the lower portion of the display will show 
p-

n

 (

n

 = memory position number 1-99). 

3. 

Press the LOG key momentarily to log a reading into memory. The REC icon will flash each 
time a data point is stored (the SCAN SD icon will appear when the meter accesses the card). 

4. 

Use the ▲ and ▼ keys to select one of the 99 data memory positions in which to record.   

5. 

To exit the manual datalogging mode, press and hold the LOG key for at least 1.5 seconds. 

Automatic Datalogging 

In automatic datalogging mode the meter takes and stores a reading at a user-specified sampling 
rate onto SD card. The meter defaults to a sampling rate of one second. To change the sampling 
rate, refer to the Setup Mode section (the sampling rate cannot be 

‘0’ for automatic datalogging): 

1. 

To begin an automatic Datalog session press and hold the LOG key for at least 1.5 seconds. 

2. 

The meter will scan for an SD card and verify that it can be used to store data. If a card is not 
inserted or if the card is defective, the meter will display SCAN SD indefinitely. In this case, 
switch the meter OFF and try again with a valid SD card. 

3. 

If the SD card is valid, the display will show the LOG icon (or the LOG icon alternating with the 
temperature display) and then the REC icon will flash each time that a reading is stored. 

4. 

To pause the datalogger, press the LOG key momentarily. The REC icon will stop flashing. To 
resume logging simply press the LOG key again momentarily. 

5. 

To terminate the datalogging session press and hold the LOG key for at least 1.5 seconds. 

6. 

When an SD card is used for the first time a folder is created on the card and named 

AMC01

Up to 99 spreadsheet documents (each with 30,000 readings) can be stored in this folder. 

7. 

When datalogging begins a new spreadsheet document named 

AMC01001.xls

 is created on 

the SD card in the AMC01 folder. The data recorded will be placed in the AMC01001.xls 
document until 30,000 readings are reached. 

8. 

If the measurement session exceeds 30,000 readings, a new document will be created 
(AMC01002.xls) where another 30,000 readings can be stored. This method continues for up 
to 99 documents, after which another folder is created (AMC02) where another 99 spreadsheet 
documents can be stored. This process continues in this same fashion with folders AMC03 
through AMC10 (last allowable folder). 

Содержание SDL300

Страница 1: ...User s Guide Vane Thermo Anemometer Datalogger Model SDL300 ...

Страница 2: ...ernational Safety Symbols This symbol adjacent to another symbol or terminal indicates the user must refer to the manual for further information Meter Description 1 Thermocouple input 2 Vane input plug 3 Air velocity reading 4 Units of measure 5 HOLD Backlight key 6 MAX MIN key 7 SET and Clock key 8 PC interface jack 9 Reset key 10 Power Adaptor jack 11 SD card slot 12 ENTER and LOG key 13 Down ar...

Страница 3: ... FPM feet per minute Km h kilometers per hour knots and mph miles per hour 1 Press and hold the UNIT key for at least 1 5 seconds to switch from one unit of measure to another 2 Note that pressing and holding the UNIT key continuously allows for quicker scrolling Release the key when the desired unit of measure is displayed Thermcouple Type J or K Temperature Mode 1 Select the Thermocouple Tempera...

Страница 4: ... the display backlight ON or OFF press and hold the backlight key for at least 1 5 seconds The meter will beep when switching the backlight ON or OFF unless the beeper is disabled as described in the Setup Mode section of this user guide System Reset If the meter s keys become inoperable or if the display freezes the Reset button can be used to reset the instrument 1 Use a paper clip or similar it...

Страница 5: ...talogging mode press and hold the LOG key for at least 1 5 seconds Automatic Datalogging In automatic datalogging mode the meter takes and stores a reading at a user specified sampling rate onto SD card The meter defaults to a sampling rate of one second To change the sampling rate refer to the Setup Mode section the sampling rate cannot be 0 for automatic datalogging 1 To begin an automatic Datal...

Страница 6: ...e PC does not have an SD card slot use an SD card adaptor available at most outlets where computer accessories are sold 4 Power the PC and run a spreadsheet software program Open the saved documents in the spreadsheet software program see examples of spreadsheet displays below Spreadsheet data example Spreadsheet example Plotting the data RS 232 USB PC Interface The optional 407001A software kit s...

Страница 7: ...t SdF SD memory card Format t CF Select the Temperature unit of measure to C or F tYPE Select the thermocouple type to K or J Setting the Clock Time 1 Access the dAtE parameter 2 Use the arrow keys to change a value 3 Use the ENTER button to step through the selections 4 Press and hold the SET button for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the met...

Страница 8: ...ch to the normal operating mode SD Card FORMATTING 1 Access the Sd F parameter 2 Use the arrow keys to select YES to format the card select NO to abort Note that all data on the card will be lost if formatting is attempted 3 Press ENTER to confirm selection 4 Press ENTER again to re confirm 5 The meter will automatically return to the normal operating mode when formatting is complete If not press ...

Страница 9: ...e rear of the meter directly above the top of the tilt stand 2 Remove and safely place the battery compartment and screws where they will not be damaged or lost 3 Replace the six 6 1 5V AA batteries observing polarity 4 Replace the battery compartment cover with the two 2 Phillips screws All EU users are legally bound by the battery ordinance to return all used batteries to collection points in yo...

Страница 10: ...atic temperature compensation for the Anemometer function and the type K J thermometer function Data Hold Freeze the display reading Memory Recall Record and Recall the Maximum and Minimum readings Display update rate Approx 1 second Data Output RS 232 USB PC computer interface for use with optional software kit part number 407001A Operating Temperature 0 to 50 C 32 to 122 F Operating Humidity 85 ...

Страница 11: ...158 F Resolution 0 1 C 0 1 F Accuracy 0 8 C 1 5 F Type K J Thermocouple Thermometer Temperature Sensor Type Resolution Range Accuracy Type K 0 1 C 50 0 to 1300 0 C 100 0 to 50 1C 0 4 0 5 C 0 4 1 C 0 1 F 58 0 to 2372 0 F 148 0 to 58 1 F 0 4 1 F 0 4 1 8 F Type J 0 1 C 50 0 to 1200 0 C 100 0 to 50 1 C 0 4 0 5 C 0 4 1 C 0 1 F 58 0 to 2192 0 F 148 0 to 58 1 F 0 4 1 F 0 4 1 8 F Note Above specifications...

Страница 12: ...is product Extech recommends that annual calibrations be performed to verify meter performance and accuracy Garantie EXTECH INSTRUMENTS CORPORATION une société FLIR garantit que cet instrument est exempt de défectuosité ou de défaut de fabrication pendant une période de trois ans à compter de la date de livraison une garantie limitée de six mois est applicable pour les sondes et les câbles S il s ...

Страница 13: ...acciones del usuario como el mal uso alambrado equivocado operación fuera de las especificaciones mantenimiento o reparación inadecuada o modificación no autorizada Extech específicamente rechaza cualesquier garantías implícitas o factibilidad de comercialización o aptitud para cualquier propósito determinado y no será responsable por cualesquier daños directos indirectos incidentales o consecuent...

Отзывы: