Congratulations on your purchase of the Extech SDL300 Thermo-Anemometer Datalogger. This 
meter displays and stores air velocity and temperature readings. Temperature readings from the 
thermometer built into the vane probe or from an externally connected type K or J thermocouple 
temperature probe can be displayed and recorded. Data is stored on an SD card for transfer to a 
PC. In addition, an RS232 port allows data streaming to a PC. This meter is shipped fully tested and 
calibrated and, with proper use, will provide years of reliable service. 

 
 

Safety 

International Safety Symbols 

This symbol, adjacent to another symbol or terminal, indicates the user must refer to the 
manual for further information. 
 

 

Meter Description 

1. 

Thermocouple input 

2. 

Vane input plug 

3. 

Air velocity reading 

4. 

Units of measure 

5. 

HOLD / Backlight 

  key 

6. 

MAX-MIN key 

7. 

SET and Clock 

  key 

8. 

PC interface jack 

9. 

Reset key 

10.  Power Adaptor jack 

11.  SD card slot 

12.  ENTER and LOG key 

13.  D

own ▼arrow and FUNCTION key 

14. 

Up ▲ arrow and UNITS key 

15.  Power ON-OFF

 key 

16.  Temperature reading with units of measure 

17.  Anemometer vane 

18.  Vane icon faces the air flow 

19.  Vane handle 

20.  Vane connecting cable 

 

Notes:

 

Battery compartment, tilt stand, and tripod mount are located on the rear of the instrument. 

Items 8, 9, and 10 are located behind the snap-off compartment cover on right side of meter. 

Содержание SDL300

Страница 1: ...User s Guide Vane Thermo Anemometer Datalogger Model SDL300 ...

Страница 2: ...ernational Safety Symbols This symbol adjacent to another symbol or terminal indicates the user must refer to the manual for further information Meter Description 1 Thermocouple input 2 Vane input plug 3 Air velocity reading 4 Units of measure 5 HOLD Backlight key 6 MAX MIN key 7 SET and Clock key 8 PC interface jack 9 Reset key 10 Power Adaptor jack 11 SD card slot 12 ENTER and LOG key 13 Down ar...

Страница 3: ... FPM feet per minute Km h kilometers per hour knots and mph miles per hour 1 Press and hold the UNIT key for at least 1 5 seconds to switch from one unit of measure to another 2 Note that pressing and holding the UNIT key continuously allows for quicker scrolling Release the key when the desired unit of measure is displayed Thermcouple Type J or K Temperature Mode 1 Select the Thermocouple Tempera...

Страница 4: ... the display backlight ON or OFF press and hold the backlight key for at least 1 5 seconds The meter will beep when switching the backlight ON or OFF unless the beeper is disabled as described in the Setup Mode section of this user guide System Reset If the meter s keys become inoperable or if the display freezes the Reset button can be used to reset the instrument 1 Use a paper clip or similar it...

Страница 5: ...talogging mode press and hold the LOG key for at least 1 5 seconds Automatic Datalogging In automatic datalogging mode the meter takes and stores a reading at a user specified sampling rate onto SD card The meter defaults to a sampling rate of one second To change the sampling rate refer to the Setup Mode section the sampling rate cannot be 0 for automatic datalogging 1 To begin an automatic Datal...

Страница 6: ...e PC does not have an SD card slot use an SD card adaptor available at most outlets where computer accessories are sold 4 Power the PC and run a spreadsheet software program Open the saved documents in the spreadsheet software program see examples of spreadsheet displays below Spreadsheet data example Spreadsheet example Plotting the data RS 232 USB PC Interface The optional 407001A software kit s...

Страница 7: ...t SdF SD memory card Format t CF Select the Temperature unit of measure to C or F tYPE Select the thermocouple type to K or J Setting the Clock Time 1 Access the dAtE parameter 2 Use the arrow keys to change a value 3 Use the ENTER button to step through the selections 4 Press and hold the SET button for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the met...

Страница 8: ...ch to the normal operating mode SD Card FORMATTING 1 Access the Sd F parameter 2 Use the arrow keys to select YES to format the card select NO to abort Note that all data on the card will be lost if formatting is attempted 3 Press ENTER to confirm selection 4 Press ENTER again to re confirm 5 The meter will automatically return to the normal operating mode when formatting is complete If not press ...

Страница 9: ...e rear of the meter directly above the top of the tilt stand 2 Remove and safely place the battery compartment and screws where they will not be damaged or lost 3 Replace the six 6 1 5V AA batteries observing polarity 4 Replace the battery compartment cover with the two 2 Phillips screws All EU users are legally bound by the battery ordinance to return all used batteries to collection points in yo...

Страница 10: ...atic temperature compensation for the Anemometer function and the type K J thermometer function Data Hold Freeze the display reading Memory Recall Record and Recall the Maximum and Minimum readings Display update rate Approx 1 second Data Output RS 232 USB PC computer interface for use with optional software kit part number 407001A Operating Temperature 0 to 50 C 32 to 122 F Operating Humidity 85 ...

Страница 11: ...158 F Resolution 0 1 C 0 1 F Accuracy 0 8 C 1 5 F Type K J Thermocouple Thermometer Temperature Sensor Type Resolution Range Accuracy Type K 0 1 C 50 0 to 1300 0 C 100 0 to 50 1C 0 4 0 5 C 0 4 1 C 0 1 F 58 0 to 2372 0 F 148 0 to 58 1 F 0 4 1 F 0 4 1 8 F Type J 0 1 C 50 0 to 1200 0 C 100 0 to 50 1 C 0 4 0 5 C 0 4 1 C 0 1 F 58 0 to 2192 0 F 148 0 to 58 1 F 0 4 1 F 0 4 1 8 F Note Above specifications...

Страница 12: ...is product Extech recommends that annual calibrations be performed to verify meter performance and accuracy Garantie EXTECH INSTRUMENTS CORPORATION une société FLIR garantit que cet instrument est exempt de défectuosité ou de défaut de fabrication pendant une période de trois ans à compter de la date de livraison une garantie limitée de six mois est applicable pour les sondes et les câbles S il s ...

Страница 13: ...acciones del usuario como el mal uso alambrado equivocado operación fuera de las especificaciones mantenimiento o reparación inadecuada o modificación no autorizada Extech específicamente rechaza cualesquier garantías implícitas o factibilidad de comercialización o aptitud para cualquier propósito determinado y no será responsable por cualesquier daños directos indirectos incidentales o consecuent...

Отзывы: