10 MC200-EU-EN V2.1 07/13
The Digital magnification can be adjusted using the zoom out ( ) and zoom in ( )
buttons. The magnification is divided into 8 stages and can be up to 4 times with 0.5
times per stage. The image can be enlarged to fifty times when combining optical
and digital magnifications. The magnification at different stages is as follows:
10x optical mag. [10* (1+0.5*stage)]
Stage
1
2
3
4
Mag. 15x 20x 25x 30x
Stage
5
6
7
8
Mag. 35x 40x 45 50x
If the optical magnification is adjusted to 40 times, the total magnification can be
200 times. Magnification at different stages is follows:
40x optical mag. [40* (1+0.5*stage)]
Stage 1
2
3
4
Mag. 60x 80x 100x 120x
Stage 5
6
7
8
Mag. 140x 160x 180x 200x
(I) Digital Magnification Stages:
0 1 2 3 4 5 6 7 8
(J) Alignment Grid : Press and hold the
button to display/hide the alignment grid. Use
the grid to assist in aligning the observed article.
(K) Cross-hair Alignment : Press and hold the
button to display/hide the
‘cross-hair’alignment tool. Use the cross-hairs to indicate the position in which the lens of
the instrument is located and to make observation alignment more convenient.
(L) Scale: Press
(Confirmation button) to display/hide the scale. Press and hold the
button to select the desired scale. After the selection, release the button. The two
available scales are shown as follows:
L-shaped Scale Crossing Scale
The display scale may be inaccurate when concave or convex materials are
observed. For standard focus, please use the flush shot method [Please refer to
item (2) of Section 1.3].
The scales vary as the magnification rate changes. The scale unit of measure is
「
μ
m
」
(micro meters). Refer to the following reference table.
Digital Magnification Scales in the 10x and 40x Optical Rates
Rate 10x 15x 20x 25x 30x 35x 40x 45x 50x
Scale 105 70.5 52.5 42.0 35.0 30.0 26.5 23.5 21.0
Rate 40x 60x 80x 100x 120x 140x 160x 180x 200x
Scale 26.5 17.6 13.2 10.6 8.88 7.60 6.52 5.76 5.32