MA1500-en-GB_v1.4   10/18 

Introduction 

Congratulations on your purchase of this MA1500 True RMS Clamp Meter. This meter measures 

AC Current, DC Current, AC/DC Voltage, Resistance, Capacitance, Frequency, Diode Test, Duty 

Cycle and Continuity. The molded case is designed for heavy duty use. This meter is shipped fully 

tested and calibrated and, with proper use, will provide years of reliable service. 

Safety 

International Safety Symbols 

This symbol, adjacent to another symbol or terminal, indicates the user must refer 

to the manual for further information. 

 

This symbol, adjacent to a terminal, indicates that, under normal use, hazardous 

voltages may be present 

 

Double insulation 

 

This 

WARNING

 symbol indicates a potentially hazardous situation, which if not 

avoided, could result in death or serious injury. 

This 

CAUTION

 symbol indicates a potentially hazardous situation, which if not 

avoided, may result damage to the product. 

 

 

 

PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY 

OVERVOLTAGE CATEGORY I 

Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which 

measures are taken to limit the transient overvoltages to an appropriate low level. 

Note – Examples include protected electronic circuits. 

OVERVOLTAGE CATEGORY II 

Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from 

the fixed installation. 

Note – Examples include household, office, and laboratory appliances. 

OVERVOLTAGE CATEGORY III 

Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations. 

Note – Examples include switches in the fixed installation and some equipment for industrial use with 

permanent connection to the fixed installation. 

OVERVOLTAGE CATEGORY IV 

Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation. 

Note – Examples include electricity meters and primary over-current protection equipment

 

WARNING 

CAUTION 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание MA1500

Страница 1: ...User Manual 1500Amp True RMS AC DC Clamp Meter Model MA1500 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...voided may result damage to the product PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY OVERVOLTAGE CATEGORY I Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measur...

Страница 3: ...discharge capacitors and remove power from the device under test before performing Diode Resistance or Continuity tests Voltage checks on electrical outlets can be difficult and misleading because of...

Страница 4: ...input jacks Display icons Description HOLD Data Hold APO Auto Power Off AUTO Autoranging P Peak Hold DC Direct Current AC Alternating Current MAX Max reading MIN Min reading Low battery ZERO DCA or CA...

Страница 5: ...ed with high sensitivity Static electricity or other sources of energy may randomly trip the sensor This is normal operation AC DC Current Measurements WARNING Disconnect the test leads before measuri...

Страница 6: ...be measured Touch the red test probe tip to the other side of the device to be measured 4 Read the resistance value in the display Continuity Test 1 Set the function switch to the position 2 Insert th...

Страница 7: ...MA1500 en GB_v1 4 10 18 7 If both readings are very small or 0 the device is shorted www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 8: ...e Hz position 2 Insert the black test lead banana plug into the negative COM jack Insert the red test lead banana plug into the positive Hz jack 3 Touch the black test probe tip to one side of the dev...

Страница 9: ...measurement situations requiring that a range be manually selected perform the following 1 Press the RANGE button The AUTO display icon will turn off 2 Press the RANGE key to step through the availabl...

Страница 10: ...crew secures the rear battery door 2 Open the battery compartment 3 Replace the 9V battery 4 Secure the battery compartment door with the screw Never dispose of used batteries or rechargeable batterie...

Страница 11: ...nge Resistance 400 00 0 5 9 digits 4 0000 k 1 0 4 digits 40 000 k 400 00 k 4 0000 M 2 0 10 digits 40 000 M 3 0 10 digits Capacitance 400 00 nF 3 5 reading 40 digits 4000 0 nF 3 5 reading 10 digits 40...

Страница 12: ...C and VAC Crest Factor 3 0 in 400A range 1 4 in 1000A range 50 60Hz and 5 to 100 of range Operating Temperature 5 C to 40 C 41 F to 104 F Storage Temperature 20 C to 60 C 4 F to 140 F Operating Humidi...

Отзывы: