480836 V5.0 8/09
6
3-2 Electrical Specifications
Unless otherwise stated, the following specifications hold under the following
conditions:
•
The meter is located in the far-field of a source, the sensor head is pointed towards
the source.
•
Ambient temperature: +23
℃
±
3
℃
•
Relative air humidity: 25% to 75%
Sensor type:
Electrical field (E)
Frequency ranges:
900MHz, 1800MHz, and 2.7GHz (measurements can be made, for
reference purposes only, using the entire range of 50MHz to 3.5GHz)
Specified measurement range:
•
CW signal (f > 900MHz)
: 20mV/m to 108.0V/m ,
53
μ
A/m to 286.4mA/m,
1
μ
W/m
2
to 30.93W/m
2
,
0
μ
W/cm
2
to 3.093mW/cm
2
Dynamic range:
Typically 75dB
Absolute error at 1 V/m and 50 MHz:
±
1.0dB
Frequency response:
•
Sensor (taking into account typical CAL factors)
:
±
1.0dB (900MHz, 1800MHz)
±
2.4dB (2.7GHz)
•
Isotropy deviation
: Typically
±
1.0dB (f>900MHz)
•
Overload limit
: 10.61mW/cm
2
(200V/m)
•
Thermal response (0 to 50
℃
)
:
±
0.2dB