background image

1                                     433201       Ver. 1.04    11/99

 

 

INSTRUCTION MANUAL 

English / Español 

Model

 

433201 and 433202 

Microprocessor Calibrator/Thermometer 

 

 

••

 Source and Measure 8 Thermocouple types

 

 

 

 

Source and Measure mV

 

 

 

••

 

O

C / 

O

F Switchable 

 

 

 

 

 

 

 

 

 

 

1. INTRODUCTION 

Congratulations on your purchase of Extech's 433201 (115V) or 433202 (220V) 

Microprocessor Thermocouple Calibrator/Thermometer. This calibrator has been 

designed with special cold junction compensation circuitry and a complex precision 

resistor network to facilitate the calibration of other thermocouple devices. They also 

provide a precision millivolt output which simulates temperature values as shown on the 

LCD display. The unique "flip-up" cover adjusts for best viewing angle and the meter 

automatically shuts off when the cover is closed. A precision millivolt signal is either 

input from a thermocouple (“Measure” mode”) or output to another thermocouple device 

(“Calibrate” mode) and the corresponding temperature, according to standard 

temperature tables, is displayed on the meter’s 4 digit LED display. 

 

2. SPECIFICATIONS 

 

Calibrate & Measure 

Resolution 

Accuracy 

Type J 

-58 to 1830

O

F (-50 to 1000

O

C) 

Type K 

-58 to 2498

O

F (-50 to 1370

O

C) 

Type T 

-184 to 752

O

F (-120 to 400

O

C) 

Type E 

-58 to 1382

O

F (-50 to 750

O

C) 

0.1

O

 to 999.9

O

 

1

O

 above 999.9

O

 

Type C 

 32 to 4172

O

F (0 to 2300

O

C) 

Type R 

 32 to 3182

O

F (0 to 1750

O

C) 

Type S 

 32 to 3182

O

F (0 to 1750

O

C) 

Type N 

-58 to 2372

O

F (-50 to 1300

O

C) 

1.0° 

±(0.15% rdg + 1.8

O

For1

O

C) 

Voltage  -5.00mV to +55.00mV  

10

µ

±(10

µ

V + 1 digit) 

Cold junction compensation         0.03

O

C/

O

C (0.02

O

F/

O

F) 

Input impedance                          10M

 

Sampling time                              4 times / second 

Display                                        4 digit LED 

Содержание 433201

Страница 1: ...ing angle and the meter automatically shuts off when the cover is closed A precision millivolt signal is either input from a thermocouple Measure mode or output to another thermocouple device Calibrate mode and the corresponding temperature according to standard temperature tables is displayed on the meter s 4 digit LED display 2 SPECIFICATIONS Calibrate Measure Resolution Accuracy Type J 58 to 18...

Страница 2: ...arge will provide one hour of continuous use Turning the meter OFF when not in use will prolong battery life 5 MEASURING TEMPERATURE 5 1 Press the red power switch to the ON position 5 2 Select the MEASURE mode on the side panel slide switch 5 3 Select the thermocouple type of your measuring probe via the faceplate TYPE button An LED will light above the T C type selected in the display Each time ...

Страница 3: ...ine battery may be used in place of the rechargeable 8 4V NiCad battery that comes with the unit 8 3 A flashing F or C on the LED display indicates a low battery condition 9 CALIBRATION REPAIR SERVICES Extech offers complete repair and calibration services for all of the products we sell For periodic calibration NIST certification or repair of any Extech product call customer service for details o...

Страница 4: ...gulo de visión y el medidor se apaga automáticamente cuando la cubierta se cierra Se introduce una señal de milivatios precisa del termopar Modo de medición o sale a otro dispositivo termopar Modo calibración y la temperatura correspondiente de conformidad con tablas estándar de temperatura se muestra in la pantalla LED de 4 dígitos del medidor 2 ESPECIFICACIONES Calibración y Medición Resolución ...

Страница 5: ... y conecte las terminales de broche en las terminales de la batería Cierre firmemente la cubierta de la batería 4 8 Cargue la batería NiCad durante 24 a 48 horas antes de su primer uso Una carga de una noche proporcionará energía para 45 minutos de operación continua aproximadamente Una carga completa proporcionará una hora de uso continuo Apagar el medidor cuando no esté en uso prolongará la vida...

Страница 6: ... al dispositivo a ser calibrado 6 17 Permita suficiente tiempo de calentamiento para el Calibrador y el Dispositivo a ser calibrado Cuando menos 30 minutos es la recomendación normal 6 18 Ajuste una salida deseada de temperatura usando los botones arriba abajo up down de ajuste de COURSE en la carátula y la perilla de ajuste fino FINE en el panel lateral La salida de temperatura será mostrada en l...

Страница 7: ...cio a Clientes al teléfono 781 890 7440 para solicitar autorización Se debe emitir un número de Autorización de Retorno RA antes de regresar cualquier producto a Extech El remitente es responsable por los cargos de envío flete seguro y empaque adecuado para prevenir daños en tránsito Esta garantía no se aplica a defectos que resulten por acciones del usuario como mal uso alambrado inapropiado oper...

Отзывы: