3                                     433201       Ver. 1.04    11/99

 

6.5  Insert the calibration cable into the sub-miniature female connector on the side panel. 

Be sure to check that it is inserted with the proper polarity and make sure the proper 

type is selected in the display. 

6.6 

Note:

 For greater accuracy, especially if there is a difference between the internal 

temperature of the calibrator and the device under calibration, we recommend that 

you use a thermocouple cable rather than the copper cable that is supplied with the 

calibrator. The thermocouple cable must match the thermocouple type of the device 

under calibration. 

6.7  Connect the other end of the cable to the device to be calibrated. 
6.8  Allow the proper warm up time for the Calibrator and the Device to be calibrated. At 

least 30 minutes is typically recommended. 

6.9  Adjust to the desired output temperature using the up/down “COURSE” adjustment 

buttons on the faceplate and the “FINE” adjustment knob on the side panel. The 

output temperature will be displayed on the LED display. The “COURSE” adjustment 

has 16 increments and will display “high” or “low” when out of range. 

7.  DIRECT MV CALIBRATION AND MEASUREMENT 

For directly calibration or measuring thermocouple devices in mV rather than in 

temperature, press the “TYPE” button until the F or C in the LED display disappears (only 

the “

O

” remains). The LED display will now indicate in mV rather than temperature units. 

 

8.  BATTERY OPERATION 

WARNING

:  DO NOT CONNECT THE AC ADAPTOR TO THE METER IF A 9V 

ALKALINE BATTERY IS INSTALLED. 

8.1  For field applications, it is recommended that the operator use the rechargeable 

NiCad battery that comes with 

8.2  A quality 9V alkaline battery may be used in place of the rechargeable (8.4V) NiCad 

battery that comes with the unit. 

8.3  A flashing “F” or “C” on the LED display indicates a low battery condition. 

 

 

9. CALIBRATION / REPAIR SERVICES 

Extech offers complete repair and calibration services

 for all of the products we sell. 

For periodic calibration, NIST certification or repair of any Extech product, call customer 

service for details on services available. Extech recommends that calibration be 

performed on an annual basis to insure calibration integrity. 

 

10. WARRANTY 

EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and 

workmanship for one year from date of shipment (a six month limited warranty applies on sensors and 

cables). If it should become necessary to return the instrument for service during or beyond the warranty 

period, contact the Customer Service Department at (781) 890-7440 for authorization. A Return 

Authorization (RA) number must be issued before any product is returned to Extech. The sender is 

responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. 

This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, 

operation outside of specification, improper maintenance or repair, or unauthorized modification. Extech 

specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not 

be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is limited to 

repair or replacement of the product. 

The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or 

implied. 

 

Copyright 

©

 

1999 Extech Instruments Corporation. All rights reserved including the 

right of reproduction in whole or in part in any form.

 

Содержание 433201

Страница 1: ...ing angle and the meter automatically shuts off when the cover is closed A precision millivolt signal is either input from a thermocouple Measure mode or output to another thermocouple device Calibrate mode and the corresponding temperature according to standard temperature tables is displayed on the meter s 4 digit LED display 2 SPECIFICATIONS Calibrate Measure Resolution Accuracy Type J 58 to 18...

Страница 2: ...arge will provide one hour of continuous use Turning the meter OFF when not in use will prolong battery life 5 MEASURING TEMPERATURE 5 1 Press the red power switch to the ON position 5 2 Select the MEASURE mode on the side panel slide switch 5 3 Select the thermocouple type of your measuring probe via the faceplate TYPE button An LED will light above the T C type selected in the display Each time ...

Страница 3: ...ine battery may be used in place of the rechargeable 8 4V NiCad battery that comes with the unit 8 3 A flashing F or C on the LED display indicates a low battery condition 9 CALIBRATION REPAIR SERVICES Extech offers complete repair and calibration services for all of the products we sell For periodic calibration NIST certification or repair of any Extech product call customer service for details o...

Страница 4: ...gulo de visión y el medidor se apaga automáticamente cuando la cubierta se cierra Se introduce una señal de milivatios precisa del termopar Modo de medición o sale a otro dispositivo termopar Modo calibración y la temperatura correspondiente de conformidad con tablas estándar de temperatura se muestra in la pantalla LED de 4 dígitos del medidor 2 ESPECIFICACIONES Calibración y Medición Resolución ...

Страница 5: ... y conecte las terminales de broche en las terminales de la batería Cierre firmemente la cubierta de la batería 4 8 Cargue la batería NiCad durante 24 a 48 horas antes de su primer uso Una carga de una noche proporcionará energía para 45 minutos de operación continua aproximadamente Una carga completa proporcionará una hora de uso continuo Apagar el medidor cuando no esté en uso prolongará la vida...

Страница 6: ... al dispositivo a ser calibrado 6 17 Permita suficiente tiempo de calentamiento para el Calibrador y el Dispositivo a ser calibrado Cuando menos 30 minutos es la recomendación normal 6 18 Ajuste una salida deseada de temperatura usando los botones arriba abajo up down de ajuste de COURSE en la carátula y la perilla de ajuste fino FINE en el panel lateral La salida de temperatura será mostrada en l...

Страница 7: ...cio a Clientes al teléfono 781 890 7440 para solicitar autorización Se debe emitir un número de Autorización de Retorno RA antes de regresar cualquier producto a Extech El remitente es responsable por los cargos de envío flete seguro y empaque adecuado para prevenir daños en tránsito Esta garantía no se aplica a defectos que resulten por acciones del usuario como mal uso alambrado inapropiado oper...

Отзывы: