OPERATING INSTRUCTIONS
Digital Thermometer
ZI-9613
www.zicotech.com
INTRODUCTION
This instrument is a portable 3
1
/
2
digit.,
compact-sized digital thermometer designed to
use external K-type thermocouple as temperature
sensor. Temperature indication follows National
Bureau
of
Standards
and
IEC584
temperature/voltage
tables
for
K-type
thermocouples. two K-type thermocouples is
supplied with the thermometer.
SAFETY INFORMATION
It is recommended that you read the safety and
operation instructions before using the thermometer.
WARNING
TO AVOID ELECTRICAL SHOCK. DO NTO
USE THIS INSTRUMENT WHEN VOLTAGES
AT THE MEASUREMENT SURFACE EXCEED
24V AC OR 60V DC.
WARNING
TO AVOID DAMAGE OR BURNS. DO NOT
MAKE TEMPERATURE MEASUREMENTS IN
MICROWAVE OVENS.
CAUTION
Repeated
sharp
flexing
can
break
the
thermocouple leads. To prolong lead life, avoid
sharp bends in the leads, especially near the
connector.
The symbol on the instrument indicates that
the operator must refer to an explanation in this
manual.
SPECIFICATIONS ELECTRICAL
Temperature Scale:
Celsius (
℃
), Fahrenheit (
℉
) user-selectble
Measurement Range:
-50
℃
to 1300
℃
, -58
℉
to 2000
℉
Resolution:
1
℃
or 1
℉
, 0.1
℃
or 0.1
℉
Accuracy:
Accuracy is specified for operating temperatures
over the range of 18
℃
to 28
℃
(64
℉
to 82
℉
), for l
year, not including thermocouple error.
±
2
℃
---------------------- -50
℃
to 0
℃
±
4
℉
---------------------- -58
℉
to 32
℉
±
(0.5% rdg+1
℃
)-------- 0
℃
to 1000
℃
±
(0.8% rdg+1
℃
)-------- 1000
℃
to 1300
℃
±
(0.5% rdg+2
℉
)-------- 32
℉
to 2000
℉
Temperature Coefficient:
0.1 times the applicable accuracy specification
per
℃
from 0
℃
to 18
℃
and 28
℃
to 50
℃
(32
℉
to 64
℉
and 82
℉
to 122
℉
).
Input protection:
60V de or 24V rms ac maximum input voltage on
any combination of input pins.
Reading Rate:
2.5 times per second.
Input Connector:
Accepts standard miniature thermocouple connectors
(flat blades spaced 7.9mm, center to center).
ENVIRONMENTAL
Ambient Operating Range:
0
℃
to 50
℃
(32
℉
to 122
℉
)
Storage Temperature:
-20
℃
to 60
℃
(-4
℉
to 140
℉
)
Relative Humidity:
0% to 80% (0
℃
to 35
℃
) (32
℉
to 95
℉
)
0% to 70% (35
℃
to 50
℃
) (95
℉
to 122
℉
)
GENERAL
Display:
3
1
/
2
digit liquid crystal display (LCD) with
maximum reading of 1999
Battery:
Standard 9V battery (NEDA 1604, IEC 6F22)
Dimensions:
162mm (H)
×
76mm(W)
×
38.5mm(D)
Weight:
210g
Supplied Probe:
4 foot type"K"thermocouple bead probe (Teflon
tape insulated).
Maximum insulation temperature 260
℃
(500
℉
)
Probe accuracy
±
2.2
℃
or
±
0.75% of reading
(Whichever is greater) from 0
°
to 800
℃
OPERATING INSTRUCTIONS
Selecting the Temperature Scale
Readings are displayed in either degrees Celsius
(
℃
), degrees Fahrenheit (
℉
).When the
thermometer is turned on, it is set to the
temperature scale that was in use when the
thermometer was last turned off. To change the
temperature scale, press the For
℃
or
℉
key.
Single-Thermocouple Temperature Measurement
The thermometer displays the temperature of the