background image

 42555    V1.1    12/2005 

Operating Instructions 

1.  Hold the meter from 0.5 to 4 inches (2 to 10cm) from the 

forehead.. 

2.  Point the meter at the center of the forehead and then pull 

and hold the trigger to turn the meter on and begin testing. 

3.  For the best results, slowly scan towards the ear. Often the 

area just above the temple provides the best measurement 
because of it’s proximity to the temporal artery. 

4.  The meter will hold and display the highest temperature 

measured. 

5.  The reading will remain in the display for approximately 6 second after the trigger is 

released. 

6. Press 

the 

o

F/

o

C button to change the measurement units. 

7. “

Hi

” or “

Lo

” will appear in the display if the temperature reading is outside the range of 

the meter. 

 

 

 

Battery Replacement 

When the low battery symbol          appears on the display, replace the meter’s 9V battery. 
The battery compartment is located on the bottom of the meter's handle. Open the 
compartment by sliding the battery compartment cover off in the direction of the arrow. 
Replace battery and re-install the battery compartment cover. 

Specifications 

Range / Resolution 

86.0 to 122.0

o

F (30.0 to 50.0

o

C) / 0.1

o

C/F 

Accuracy ± 

0.9

o

F (0.5

 o

C). 

Note: Accuracy is specified for the following ambient (room) 
temperature range: 73 to 82°F (23 to 27°C) 

Display  

LCD display 

 

Response time 

500ms 

Operating Humidity 

Max. 80% RH 

Power Supply 

9V battery 

Automatic Power Off 

Meter shuts off automatically after 6 seconds 

Weight       

3.6 oz. / 100g 

Dimensions 

6.7 x 1.7 x 1.6” (170 x 44 x 40mm) 

 

Содержание 42555

Страница 1: ...roduction This thermometer is intended for scanning groups of individuals or monitoring an individual for elevated temperatures It is not a substitute for a clinical thermometer Always use a clinical thermometer when high accuracy body temperature measurements are required ...

Страница 2: ... the meter in water Avoid touching and or scratching the infrared sensor lens Always use a clinical thermometer to verify any abnormal temperature measurements Clean the lens area by gently blowing with compressed air and use a damp swap to wipe the lens Do not use any solvents to clean the lens Measurement Notes 1 If the meter has been stored in a cold or hot environment allow it at least 20 minu...

Страница 3: ...ill appear in the display if the temperature reading is outside the range of the meter Battery Replacement When the low battery symbol appears on the display replace the meter s 9V battery The battery compartment is located on the bottom of the meter s handle Open the compartment by sliding the battery compartment cover off in the direction of the arrow Replace battery and re install the battery c...

Страница 4: ... before any product is returned to Extech The sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied w...

Отзывы: