background image

Version 1.5   9/02 

7

RS-232 PC Communications Interface 

RS-232 data format 

Baud rate 

9600 

Parity  

none 

Data bits 

Stop bits  

Flow control 

none 

RS-232 Connection 

Connect the supplied communications cable to the rear of the 
power supply and to the communications port of the PC. 
 
Power Supply 

 

PC 

Pin 2 ------------------------------ Pin 2 RX 
Pin 3 ------------------------------ Pin 3 TX 
Pin 4 ------------------------------ Pin 4 DTR 
Pin 5 ------------------------------ Pin 5 GND 

RS-232 Commands 

The command format is as follows: Command Parameter <Carriage Return (cr)>  
 
For example: V 20.5 (sets output to 20.5VDC) 
An ASCII code for <Carriage Return> must be sent together with the command or 
commands. Commands will not be processed until a <Carriage Return> is received. Note 
that the unit can process up to 50 characters. If more than 50 characters are received, the 
unit will clear the entire command buffer. 

Summary of Commands 

STBY <cr> 

(place the unit in standby mode) 

OPER <cr> 

(place the unit in operating mode) 

V xx.xxx <cr> 

(set the voltage output) 

A x.xxx <cr> 

(set the current output) 

SEC xxx.xx <cr>  (enter time in seconds) 
OVP xx.xxx <cr>  (enter desired voltage value for over-voltage protection) 
OCP 0 <cr> 

(disable the over-current protection) 

OCP 1 <cr> 

(enable the over-current protection) 

STORE xxx <cr>  (store the present output values to memory location xxx) 
RECALL xxx <cr>  (recall an output setup from memory location xxx) 
STEP <cr> 

(enable the ATS Auto Test Sequencing function) 

? <cr>   

(retrieve all data) 

V? <cr>   

(retrieve the voltage value) 

A? <cr>   

(retrieve the current value) 

OCP? <cr> 

(retrieve the status of OCP) 

OVP? <cr>  

(retrieve the voltage value for OVP) 

SEC? <cr> 

(retrieve the time value in seconds) 

STATUS? <cr> 

(retrieve the STBY/OPER, OCP, OVP, & 5V/3.3V fixed output status) 

 

Note

: All commands are case insensitive. 

5

3

1

9

7

Содержание 382280

Страница 1: ...382280 Programmable DC Power Supply This 200 watt power supply offers an adjustable 5ADC 40VDC output plus fixed outputs 5VDC and 3 3VDC Programming features include custom over voltage current protec...

Страница 2: ...ble Currrent 0 2 9mA Fixed 5V 0 25V Fixed 3 3V 0 16V Response Time 150 ms Temperature Coefficient 100ppm o C Load Regulation 1mV Amp Line Regulation No effect if line voltage is within 90V and 130V fo...

Страница 3: ...mming value 12 RIGHT Arrow Press to move the cursor to a new location in a row 13 POWER ON OFF Press to turn the supply on or off 14 Rubber feet 15 Dot matrix display 16 GND Ground terminal connected...

Страница 4: ...e as desired 3 To change the values of voltage or current use the V A button to move the underline cursor to first digit of the voltage or current display enter the value directly from the numeric key...

Страница 5: ...the middle of the lower line of the display Over Current Protection OCP Over Current Protection OCP allows the user to set a maximum permissible current If the current limit is exceeded the unit autom...

Страница 6: ...he first and last steps Each test setup memory location represents one step The duration the step is programmable from 0 01 to 60 seconds and is programmed when the steps are stored Note Initial setup...

Страница 7: ...ived the unit will clear the entire command buffer Summary of Commands STBY cr place the unit in standby mode OPER cr place the unit in operating mode V xx xxx cr set the voltage output A x xxx cr set...

Страница 8: ...racter V or A is the programmed value while the second set of data is the read back A D value The third set of data is the status CV CC of the unit If the third set of data shows 0V the unit is placed...

Страница 9: ...transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modif...

Отзывы: