Automated Testing 

Storing Test Setups 

1.  Set the Voltage and Current values as desired. 
2.  Press the SEC button and set the time in seconds. This time will be associated with the 

step if the automatic step output function is selected. 

3.  Press the STORE button to save the values. “

STRxxx?”

 will appear in the display 

(where “xxx” (0 to 199) will be the memory location). 

4.  Press ENTER to store the value in the displayed location or enter a new location 

number from the keypad and press ENTER. If a three digit number is entered, pressing 
ENTER is not necessary. 

Recalling Test Setups 

1.  Press the RECALL button to view the setup in a memory location. “

RCLxxx”

 and the 

stored values will appear in the display (where “xxx” (0 tp 199) will be the memory 
location). 

2.  To view the setup in any other location, enter the setup number and press ENTER. If a 

three digit number is entered, pressing ENTER is not necessary. 

3.  Press RECALL to exit the display. 

Automatic Test Sequencing (ATS) 

In ATS mode, the unit automatically cycles through a series of stored test setups. The first 
and last steps are programmable and the sequence will include all steps between the first 
and last steps. Each test setup memory location represents one step. The duration the step 
is programmable from 0.01 to 60 seconds and is programmed when the steps are stored. 

Note:

 Initial setup and sequencing should be performed in the STBY mode to verify proper 

operation without an active output. 

Setup and Operation of Automatic Test Sequencing (ATS) 

1.  To set the LAST and FIRST step of the sequence: 

a. 

Press the RECALL button. 

b. 

Use the numeric keypad to enter the three digit test setup location number for the 
last step (RCL018, for example) and press ENTER. 

c. 

Use the numeric keypad to enter the three digit test setup location number for the 
first step (RCL001, for example) and press ENTER. 

d. 

Press RECALL  

2.  The time duration for each step is stored with each step in the storing process. 
3.  To begin the test sequencing, press and hold the RECALL until the beeper sounds and 

the stair-step icon appears. The sequencing will automatically begin a single cycle 
through the programmed steps. 

4.  Observe the output for proper operation. 
5.  Press the OPER/STBY key to activate the output 
6.  Press the OPER/STBY to return to the standby mode or press CE to exit the sequencing 

mode 

Note

: The ATS session will be terminated if any command is received via the RS-232c 

interface.  

Note

: For continuous cycling, hold the RECALL button while turning power on. 

 

Содержание 382280

Страница 1: ...382280 Programmable DC Power Supply This 200 watt power supply offers an adjustable 5ADC 40VDC output plus fixed outputs 5VDC and 3 3VDC Programming features include custom over voltage current protec...

Страница 2: ...ble Currrent 0 2 9mA Fixed 5V 0 25V Fixed 3 3V 0 16V Response Time 150 ms Temperature Coefficient 100ppm o C Load Regulation 1mV Amp Line Regulation No effect if line voltage is within 90V and 130V fo...

Страница 3: ...mming value 12 RIGHT Arrow Press to move the cursor to a new location in a row 13 POWER ON OFF Press to turn the supply on or off 14 Rubber feet 15 Dot matrix display 16 GND Ground terminal connected...

Страница 4: ...e as desired 3 To change the values of voltage or current use the V A button to move the underline cursor to first digit of the voltage or current display enter the value directly from the numeric key...

Страница 5: ...the middle of the lower line of the display Over Current Protection OCP Over Current Protection OCP allows the user to set a maximum permissible current If the current limit is exceeded the unit autom...

Страница 6: ...he first and last steps Each test setup memory location represents one step The duration the step is programmable from 0 01 to 60 seconds and is programmed when the steps are stored Note Initial setup...

Страница 7: ...ived the unit will clear the entire command buffer Summary of Commands STBY cr place the unit in standby mode OPER cr place the unit in operating mode V xx xxx cr set the voltage output A x xxx cr set...

Страница 8: ...racter V or A is the programmed value while the second set of data is the read back A D value The third set of data is the status CV CC of the unit If the third set of data shows 0V the unit is placed...

Страница 9: ...transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modif...

Отзывы: