Model 382068 V3.0 4/2005
20
Datalogging
Single mode
1. Single mode datalogging records one reading at a time.
2. To record one reading, press the RECORD key until one beep heard. (Note that if the
button is held longer, 2 beeps will sound and the meter will be in continuous mode; see
below). The record number (1 through 25 or FULL) will briefly appear on the LCD.
3. The meter can store up to 25 readings. When the memory is full, FULL will display.
4. To view the readings, press and hold the READ button until two tones are heard. The
second tone is longer and louder than the first.
5. Now use the yellow and keys to scroll through the recorded readings.
Continuous mode (for use with PC interface only)
1. In continuous mode the meter records one reading after another automatically. Up to
4000 records can be recorded.
2. Press and hold the RECORD until two tones are heard (not available in 3
φ
3W and 3
φ
4W function positions). The meter will begin recording data and the RECORD icon will
appear on the LCD.
3. To exit the Record mode, press and hold the RECORD button until the RECORD
display icon disappears.
4. Readings can only be transferred to a PC with optional Windows®
software and
interface cable. Instructions for use are provided with optional software/hardware kit.
Clearing Datalog Data
1. Turn the meter off and hold down the RECORD button while turning the meter on.
2. Release the RECORD key when CLR appears. The datalog memory is now clear.
www.burntec.com