6.0
Float Test Procedure (“8x256” and “20x48” Modes)
6.1 Step1: Select DLV40 “8x256” or “20x48” Mode & Clear Old String Data
6.1.1 Option 1: Using Winmeter 5.0 Software (Recommended):
(v)
Connect the DLV40 to Winmeter 5.0 software via USB
(vi)
Once connected, the DLV40 will display
“DLV40 <-> PC”
and download data in its current
mode .
If the DLV40 is
NOT
in the required mode, the User can change to “8x256” or “20x48” mode by
the selecting the appropriate radio button.
If the Winmeter 5.0 AutoStart software is not enabled,
RUN
the
Winmeter 5.0 software to initiate data transfer
of the current DLV40 mode
(“8 by 256”, “20 by 48” or “Discharge”).
Change mode of
operation .
(If required)
If the “DLV40 & DMA35 USB Interface” does not initiate automatically
select
“Download Device” -> “Via USB (search for device)”.