SmartR™ Features
156
MAX-635G
Pair Detective
Isolation
results also display a five-level indicator of the measured
isolation/leakage resistance.
The
Summary
page allows you to set the following
parameters:
Isolation Soak Time
sets a soak period of time
from 1 to 60 seconds for the
Isolation
test. Default
is 5 s. If your unit has the HIVOLT (high voltage)
software option configured, the
Soak Voltage
parameter values are extended up to 500 V
maximum. Default is inherited from
Multimeter
Isolation
test.
Note:
The application of higher voltage over time tends to
dry-up moisture. Some faults may appear to disappear
after a soak period, but only the moisture has disappeared. The fault
remains and moisture will likely return at a later time. Therefore, do not
extend the soak time unless testing insulation resistivity (PIC, paper, etc.).
Cable Setup
button opens a new page that allows you to set the
Cable
Parameters
for the test.
Threshold Settings
button opens a new page that allows you to set the
threshold parameters. They control when
Pair Detective
identifies a
fault, such as 4 VDC for a
battery
fault. You can also modify when
Pair
Detective
identifies a high-resistive versus normal fault by changing the
threshold parameters.
Note:
Individual test parameters can be modified by accessing their respective
test pages.
To select the parameter values:
1.
Press the up/down arrow keys to highlight the desired parameter.
2.
Press the left/right arrow keys to view and select the options.
OR
Содержание MAX-635G
Страница 1: ...User Guide Copper and G fast VDSL2 ADSL2 Multi play Test Set MAX 635G www EXFO com ...
Страница 10: ......
Страница 16: ...Introducing the MAX 635G 6 MAX 635G Using the MaxTester Back Single battery Battery Door Screws Screws Stylus ...
Страница 26: ......
Страница 162: ......
Страница 218: ......
Страница 270: ...TDR 260 MAX 635G Continuous Examples of Common Faults Short Fault ...
Страница 271: ...TDR Copper and G fast VDSL2 ADSL2 Multi play Test Set 261 Continuous Open ...
Страница 272: ...TDR 262 MAX 635G Continuous Bridged Tap ...
Страница 273: ...TDR Copper and G fast VDSL2 ADSL2 Multi play Test Set 263 Continuous Splice Joint ...
Страница 274: ...TDR 264 MAX 635G Continuous Peak Function ...
Страница 275: ...TDR Copper and G fast VDSL2 ADSL2 Multi play Test Set 265 Continuous Dual Trace View ...
Страница 276: ...TDR 266 MAX 635G Continuous Load Coil ...
Страница 290: ......
Страница 336: ......
Страница 414: ......
Страница 424: ......
Страница 432: ......
Страница 438: ......
Страница 444: ......