SCPI Command Reference
Variable Attenuator
119
Product-Specific Commands—Description
:OUTPut[1..n]:ALC[:STATe]
Description
This command activates or deactivates the
power tracking that controls the output power
level. The power tracking is done via the ALC
(Automatic Leveling Control) loop. The state of
the ALC loop (on or off) is used only when the
POWer control mode is active.
At *RST, this value is set to off.
Syntax
:OUTPut[1..n]:ALC[:STATe]<wsp><State>
Parameter(s)
State:
The program data syntax for <State> is defined
as a <Boolean Program Data> element. The
<State> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
The <State> parameter corresponds to the new
state of the ALC (Automatic Leveling Control)
loop.
OFF: No power tracking will be performed.
ON: Power tracking is activated.
See Also
:OUTPut[1..n]:ALC[:STATe]?
:CONTrol[1..n]:MODE
:OUTPut[1..n]:DTOlerance
:OUTPut[1..n]:POWer
:OUTPut[1..n]:RPOWer
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание IQS-3150
Страница 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 29: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 55: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 75: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 87: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...